20 | Keysight | E4990A Impedance Analyzer 20 Hz to 10/20/30/50/120 MHz - Data Sheet
Display Function
Display
Size/type
Number or pixels1
10.4 inch color LCD (TFT)
1024 X 768 (XGA)
Scale type
X axis scale
Linear and log
Y axis scale
Linear and log (depends on the sweep type)
Number of traces
Data trace
4 traces per channel
Memory trace
Data math function
4 traces per channel
Data + memory, data - memory, data × memory, data/memory, offset, equation editor
1. Valid pixels are 99.99% and more. Below 0.01% of fixed points of black, blue, green or red are not regarded as failure.
Marker function
Marker type and number
10 independent markers per trace. Reference marker available for delta marker operation.
Marker search
Search type
Maximum value, minimum value, multi-peak, multi-target, peak, peak left, peak right, target, target left, target
right, and width parameters with user defined bandwidth values.
Search track
Search range
Marker X-axis display
Others
Performs search by each sweep
User define
Sweep parameter value, sweep elapsed time, or relaxation time (1/( 2πf))
Marker continuous mode, ∆ marker mode, marker coupled mode,
Marker value substitution (Marker→), marker zooming, marker table, marker statistics
Equivalent circuit analysis
Circuit model
3-component model (4 models), 4-component model (3 models)
Analysis type
Equivalent circuit parameters calculation, frequency characteristics simulation
Limit Line Test
Funcions
Define the test limit lines that appear on the display for pass/fail testing. Defined limits may be any combina-
tion of horizontal/sloping lines and discrete data points.
Other functions
Beep fail, Limit line offset