32 | Keysight | E4980A Precision LCR Meter 20 Hz to 2 MHz, E4980AL Precision LCR Meter 20 Hz to 300 kHz/500 kHz/1 MHz – Data Sheet
Table 55. E4980A-005 measurement time [ms] (DC bias: OFF)
Measurement
time mode
Test frequency
20 Hz
1190
1150
1040
100 Hz 1 kHz
10 kHz 100 kHz 1 MHz
2 MHz
570
180
1
2
3
LONG
MED
650
380
240
590
200
37
580
180
25
570
180
23
570
180
23
SHORT
23
10
1
1. LONG
2. MED
3. SHORT
0.1
0.01
0.001
20
100
1k
10k
Test frequency [Hz]
100k
1M 2M
Figure 10. Measurement time (DC bias: OFF, E4980A-005)
Table 56. E4980AL measurement time [ms]
Measurement
time mode
Test frequency
20 Hz
729
100 Hz
1 kHz
10 kHz
100 kHz
343
1 MHz
1
2
3
LONG
MED
423
250
149
363
140
26
353
122
14
343
118
12
650
119
SHORT
579
12
10
1
1. LONG
2. MED
3. SHORT
0.1
0.01
0.001
20
100
1k
10k
Test frequency [Hz]
100k
1M 2M
Figure 11. Measurement time (E4980AL)