04 | Keysight | E36100 Series Programmable DC Power Supplies - Data Sheet
Performance Specifications
Tolerance %
DC output rating (0 to 40 °C)
Max. voltage
E36102A
E36103A
E36104A
E36105A
E36106A
6 V
5 A
20 V
2 A
35 V
1 A
60 V
100 V
0.4 A
Max. current
0.6 A
Load regulation ± (% of output + offset)
Voltage
<0.01% +
<0.02% +
2 mV
3 mV
6 mV
10 mV
30 µA
20 mV
20 µA
Current
250 µA
100 µA
50 µA
Line regulation ± (% of output + offset)
Voltage
<0.01% +
<0.02% +
1 mV
2 mV
4 mV
7 mV
12 mV
20 µA
Current
250 µA
100 µA
50 µA
30 µA
Output ripple and noise (20 Hz to 20 MHz)
Voltage
RMS
350 µV
10 mV
2 mV
4 mV
5 mV
15 mV
Pk-Pk
30 mV
60 mV
100 mV
150 mV
Accuracy 12 months (23 °C ± 5 °C)
Programming accuracy ± (% of output + offset)
Voltage
0.05% +
0.05% +
3 mV
5 mA
8 mV
1 mA
12 mV
20 mV
40 mV
Current
0.6 mA
0.4 mA
0.3 mA
Readback accuracy ± (% of output + offset)
Voltage
0.05% +
0.05% +
0.25% +
3 mV
4 mA
5 mV
1 mA
8 mV
12 mV
20 mV
Current
0.5 mA
0.3 mA
0.2 mA
Small Current
40 µA
40 µA
40 µA
40 µA
40 µA
(0-20 mA)
(0-8 mA)
(0-4 mA)
(0-3 mA)
(0-2 mA)
Load transient recovery time
(Time to recover to within the settling band following a load change from 50% to 100% and from 100% to 50% of full load)
Voltage settling band
Time
15 mV
50 mV
87.5 mV
150 mV
250 mV
<50 μs
<50 μs
<50 μs