欢迎访问ic37.com |
会员登录 免费注册
发布采购

DSOZ634A 参数 Datasheet PDF下载

DSOZ634A图片预览
型号: DSOZ634A
PDF下载: 下载PDF文件 查看货源
内容描述: [Infiniium Z-Series Oscilloscopes]
分类和应用:
文件页数/大小: 41 页 / 4195 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号DSOZ634A的Datasheet PDF文件第12页浏览型号DSOZ634A的Datasheet PDF文件第13页浏览型号DSOZ634A的Datasheet PDF文件第14页浏览型号DSOZ634A的Datasheet PDF文件第15页浏览型号DSOZ634A的Datasheet PDF文件第17页浏览型号DSOZ634A的Datasheet PDF文件第18页浏览型号DSOZ634A的Datasheet PDF文件第19页浏览型号DSOZ634A的Datasheet PDF文件第20页  
16 | Keysight | Infiniium Z-Series Oscilloscopes - Data Sheet  
Analysis Tools: PrecisionProbe Basic and Advanced (N2809A and N2807B)  
Now every probe and cable in the system can have the exact same response — probe to  
probe or cable to cable — without the inaccuracies that using one model can produce.  
You can properly characterize custom probes and remove unwanted responses. In  
addition to characterizing the cables, PrecisionProbe allows for immediate use on the  
Turn your Z-Series oscilloscope  
into a time-domain  
transmissometry (TDT). Quickly  
characterize and compensate  
any input into your scope.  
same instrument. PrecisionProbe saves you time and money while increasing your  
measurement accuracy.  
PrecisionProbe technology turns  
When you combine InfiniiMax probes with switches between the amplifier and the probe  
head, PrecisionProbe allows for full correction and automation of each probe’s path.  
Full automation is then available to allow for quick swapping of the inputs via Infiniium’s  
compliance framework. For increased accuracy, purchase PrecisionProbe Advanced for  
faster edge speeds and true differential measurements.  
your oscilloscope into the ultimate  
characterization tool. Not only can you  
do the normal waveform transformations  
such as de-embedding through  
InfiniiSim, PrecisionProbe allows quick  
characterization of your entire probe  
system (including cables and switches)  
without the need for extra equipment.  
PrecisionProbe takes advantage of the fast  
“cal output” signal built into the Z-Series  
to characterize and compensate insertion  
loss on the measurement system.  
PrecisionProbe technology:  
Properly creates custom probe  
transfer function = VOut/VIn  
Properly characterizes probed system  
transfer function such that  
VOut/VIn = VOut/VSrc  
Removes unwanted S21 cable loss  
PCI Express measurement comparisons  
Root complex device  
2.5 GT/s 12 GHz  
Eye height (mV)  
517.19  
312.22  
341.1  
306.6  
344.4  
96.83  
100.16  
96.92  
Eye height PrecisionProbe Gain (%)  
553.94  
348.19  
376  
348.33  
374.41  
103.09  
108.33  
106.01  
108.24  
7.1  
5 GT/s_12 GHz_3.5 dB  
5 GT/s_12 GHz_6 dB  
5 GT/s_16 GHz_3.5 dB  
5 GT/s_16 GHz_6 dB  
8 GT/s_12 GHz_P7  
8 GT/s_12 GHz_P8  
8 GT/s_16 GHz_P7  
8 GT/s_16 GHz_P8  
11.5  
10.2  
13.6  
8.7  
6.5  
8.2  
9.4  
Every Z-Series oscilloscope has Keysight’s  
custom indium phosphide (InP) fast edge.  
100.24  
8.0  
By characterizing and compensating for cable loss on the cable connected to the PCI  
Express test fixture, the designer was able to gain between 6.5 and 13.6% margin that  
would have been lost otherwise.  
 复制成功!