欢迎访问ic37.com |
会员登录 免费注册
发布采购

DSOX6B25T602BW 参数 Datasheet PDF下载

DSOX6B25T602BW图片预览
型号: DSOX6B25T602BW
PDF下载: 下载PDF文件 查看货源
内容描述: [InfiniiVision 6000 X-Series Oscilloscopes]
分类和应用:
文件页数/大小: 45 页 / 3290 K
品牌: KEYSIGHT [ Keysight Technologies ]
 浏览型号DSOX6B25T602BW的Datasheet PDF文件第1页浏览型号DSOX6B25T602BW的Datasheet PDF文件第2页浏览型号DSOX6B25T602BW的Datasheet PDF文件第3页浏览型号DSOX6B25T602BW的Datasheet PDF文件第4页浏览型号DSOX6B25T602BW的Datasheet PDF文件第6页浏览型号DSOX6B25T602BW的Datasheet PDF文件第7页浏览型号DSOX6B25T602BW的Datasheet PDF文件第8页浏览型号DSOX6B25T602BW的Datasheet PDF文件第9页  
05 | Keysight | InfiniiVision 6000 X-Series Oscilloscopes - Data Sheet  
New Standard for Price Performance: Bandwidth, Visualization, and Integration  
Bandwidth  
Superior signal integrity with total-cost- of-  
ownership leadership 6 GHz, 20 GSa/s  
When you choose your next oscilloscope, bandwidth is the most  
important specification to consider, as it defines the maximum  
frequency content your oscilloscope can acquire. Acquiring  
signals with faster edge rates or faster fundamental frequencies  
requires higher-bandwidth scopes to make the most accurate  
measurements. However, the higher the bandwidth of your  
oscilloscope, the higher the price is likely to be.  
Sample rate is the second important specification, as it  
determines the time span between each acquired sample point,  
and it ultimately becomes the limiting factor of the oscilloscope’s  
Figure 2. An oscilloscope with 1.5-GHz bandwidth captures only the  
bandwidth. In a modern oscilloscope with Brickwall filter  
response, the sample rate must be at least 2.5 times higher the  
bandwidth. So a scope with 6-GHz bandwidth requires a sample  
rate of at least 15 GSa/s to avoid aliasing.  
fundamental frequency of a 2.5-Gbps PRBS signal.  
With the InfiniiVision 6000 X-Series, you can get up to 6-GHz  
bandwidth and a 20-GSa/s sampling rate so you can confidently  
measure signals with rise times faster than 150 ps or signals with  
higher than 2-Gbps NRZ (non-return to zero) data signal rates.  
Explore Figures 1 through 4 to see the power extra bandwidth  
delivers to your measurements.  
Figure 3. An oscilloscope with 3-GHz bandwidth sees some of the 3rd  
harmonic of a 2.5-Gbps PRBS signal.  
Figure 1. Measuring the rise time of a 130-ps rise-time edge (10 to 90%).  
The rise time measurement by  
– Channel 1 at 6-GHz bandwidth (yellow): 132 ps  
– Reference 1 (R1) at 3-GHz bandwidth limit: 196 ps  
– Reference 2 (R2) at 1.5-GHz bandwidth limit: 216 ps  
Figure 4. An oscilloscope with 6-GHz bandwidth sees up to the 5th  
harmonic of the 2.5-Gbps PRBS signal. You see the true signal integrity  
of your waveform.