27 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
Horizontal system analog channels
4022A
4024A
4032A
4034A
4052A
4054A
4104A
4154A
Time base range
Time base accuracy 1
Time base delay time range
2 ns/div to 50 s/div
1 ns/div to 50 s/div
500 ps/div to 50 s/div
± 10 ppm
Pre-trigger
Greater of 1 screen width or 200 μs (400 μs in interleaving mode)
Post-trigger 1 to 500 s
± 100 ns
Channel-to-channel deskew range
Δ Time accuracy (using cursors)
± 0.001% of reading ± 0.16% screen width ± 30pS
Modes
XY
Main, zoom, roll, XY
On channels 1 and 2 only. Z Blanking on Ext Trigger Input, 1.4 V threshold
Bandwidth: Maximum bandwidth. Phase error at 1 MHz: < 0.5 degree
Time base: 200 ns/div to 50 ms/div
Horizontal system digital channels
Minimum detectable pulse width
Channel-to-channel skew
2 ns
2 ns (typical); 3 ns (maximum)
Acquisition system
4022A
4024A
4032A
4034A
4052A
4054A
4104A
4154A
Maximum analog channels sample rate
Analog channels equivalent sample rate
Maximum analog channels record length
Maximum digital channels sample rate
Maximum digital channels record length
5 GSa/s half channel interleaved, 2.5 GSa/s all channels
N/A
4 Mpts half channel interleaved, 2 Mpts all channel
1.25 GSa/s
2 Mpts (with digital channels only)
Default mode
128 Gsa/s
Modes
Normal
Peak detect
Averaging
High
Capture glitches as narrow as 200 ps at all time base settings
Selectable from 2, 4, 8, 16, 64, ... to 65,536
Real-time boxcar averaging reduces random noise and effectively increases vertical resolution
– 12 bits: ≥ 50 μs/div
resolution
– 11 bits: ≥ 20 μs/div
– 10 bits: ≥ 10 us/div
– 9 bits: ≥ 5 us/div
Segmented
Roll
Segmented memory optimizes available memory for data streams that have long dead times
between activity. Maximum segments = 1000. Re-arm time = 1 μs (minimum time between trigger
events). Re-arm time when used with the zone touch trigger = 65 μs or faster (typical)
Displays the waveform moving across the screen from right to left. Available at the time base
50 ms/div or slower
Equivalent
time
1 GHz and 1.5 GHz models only. 7.8 ps fine interpolator resolution yields a maximum effective
sample rate of 128 GSa/s
1� Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.