10 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet  
Oscilloscope Experience Redefined:  
Experience the Usability (Continued)  
Zone touch trigger  
One of the biggest challenges of using an oscilloscope is setting  
up an advanced trigger to isolate a signal of interest. While  
advanced triggers are powerful features, zone touch trigger  
provides a turnkey trigger solution.  
You simply observe the signal of interest on the display, and  
draw a zone (box) around it. What used to be hours of work can  
now take just a few seconds. If you want to move your zones to  
another location, just drag them over. The 4000 X-Series can be  
set up to easily trigger on one or two zone boxes simultaneously  
with either must intersect or must not intersect conditions.  
Zone triggering does not compromise the waveform update  
rate; the 4000 X-Series will still maintain an ultra-fast 200,000  
waveforms per second or more, even with additional features  
turned on. In other words, the oscilloscope that sees everything  
can easily trigger on anything.  
Zone touch trigger + segment memory: A whole  
new experience  
The combination of the industry’s only hardware-based zone  
touch trigger with the 4000 X-Series’ segment memory simplifies  
and enhances your debugging experience. In Figure 16, the 4000  
X-Series has isolated and captured 1,000 metastable signals,  
showing the critical bit errors over a 32-second time span at  
5 GS/s sampling rate in the segment memory. The segment  
memory also allows you to overlay all segments to identify the  
worst-case signal.  
Figure 15. When you see anomalies, all you have to do is  
draw a zone box to trigger on them.  
Figure 16. Combination of the zone touch trigger and  
segmented memory.  