28 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet  
Oscilloscope Experience Redefined:  
InfiniiVision 4000 X-Series Performance Characteristics (Continued)  
Trigger type selections  
CXPI (optional)  
Trigger on the start of frame (SOF), the end of frame (EOF), PTYPE, frame ID, data and info frame ID, data and info  
frame ID (long frame), CRC field error, parity error, inter-byte space error, inter-frame space error, framing error,  
data length error, sample error, all errors, sleep frame, wakeup pulse  
MIL-STD1553 (Option)  
SPI (Option)  
Trigger on MIL-STD 1553 signals on data word start/stop, command/status start/stop, RTA, RTA + 11 bits, and  
error conditions (parity, sync, Manchester)  
Trigger on SPI (serial peripheral interface) data pattern during a specific framing period. Supports positive and  
negative chip select framing as well as clock Idle framing and user-specified number of bits per frame. Supports  
MOSI and MISO data  
UART/RS232/422/485 (Option)  
USB (Option)  
Trigger on Rx or Tx start bit, stop bit, data content, or parity error  
Trigger on start of packet (SOP), end of packet (EOP), suspend 3, resume 3, reset 3, packets (token, data,  
handshake, or special), and errors (PID, CRC5, CRC16, glitch, bit stuff 3, SE1 3). Supports USB 2.0 low speed, full  
speed, and hi-speed (hi-speed is supported on 1 GHz and 1.5 GHz models only)  
SENT (optional)  
Trigger and decode on SENT bus. start of fast channel message, start of slow channel message, fast channel SC  
and data, slow channel message ID, slow channel message ID and data, tolerance violation, fast channel CRC  
error, slow channel CRC error, all CRC errors, pulse period error, successive sync pulses error (1/64)  
Waveform measurements and cursors  
DC vertical accuracy/cursors 2  
Single cursor accuracy: ± [DC vertical gain accuracy + DC vertical offset accuracy + 0.21% full  
scale]  
Dual cursor accuracy: + [DC vertical gain accuracy + 0.42% full scale] 1  
2 pairs of XY cursors  
Cursors  
Automatic measurement of positions, ∆X, 1/∆X, ∆Y, and ∆Y/∆X  
Measurements continuously updated with statistics. Cursors track last selected measurement.  
Select up to four measurements from the list below:  
Automatic measurements  
Voltage  
Time  
Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot, average- N  
cycles, average- full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N cycles, AC  
RMS- full screen (standard deviation), ratio (RMS1/RMS2)  
Period, frequency, counter, + width, - width, burst width, duty cycle, bit rate, rise time, fall time,  
delay, phase, X at min Y, X at max Y  
Count  
Mixed  
Positive pulse count, negative pulse count, rising edge count, falling edge count  
Area- N cycles, area- full screen  
Automatic measurement logging  
Counter  
Available via BenchVue  
Built-in frequency counter  
Source  
Any analog or digital channel  
Resolution  
5 digits. Up to 8 digits with an external reference 10 MHz input  
Max frequency Bandwidth of oscilloscope  
Standard mask/limit test capability that provides easy, fast pass/fail comparison of a signal  
Mask limit test option  
under test to a predefined mask template or auto-mask template. Predefined mask templates or  
edits to an auto-mask template can be made via a simple text editor. > 270,000 mask tests per  
second (waveform update rate)  
Waveform math  
Number of math functions  
Arithmetic  
Four (display one at a time)  
Add, subtract, multiply, divide, differentiate, integrate, FFT, Ax + B, squared, square root, absolute, common log,  
natural log, exponential, base 10 exponential, LP filter, HP filter, averaged value, smoothing, envelope, magnify,  
max hold, min hold, measurement trend, chart logic bus (timing or state)  
FFT  
Up to 64 kpts resolution. FFT window types: Hanning, flat top, rectangular, Blackman-Harris  
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration  
temperature.  
2. 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity  
setting.  
3. Suspend, resume, reset, bit stuff error, and SE1 error are USB 2.0 low and full speed only.  