24 | Keysight | Infiniium 90000 Series Oscilloscopes - Data Sheet  
Infiniium 90000A Series Oscilloscopes  
Performance characteristics  
Vertical  
Input channels  
Analog bandwidth (–3 dB) 1, 5  
Four  
90254A  
2.5 GHz  
90404A  
90604A  
90804A  
91204A  
91304A  
4 GHz  
6 GHz  
8 GHz  
12 GHz  
12 GHz  
DSP enhanced bandwidth 4  
Rise time/fall time 6  
10 to 90%  
20 to 80%  
Input impedance 7  
Sensitivity 2  
91304A: 13-GHz real-time, user-selectable DSP enhanced bandwidth  
90254A  
140 ps  
105 ps  
50 Ω, ± 3%  
1 mV/div to 1 V/div  
DC  
90404A  
105 ps  
79 ps  
90604A  
70 ps  
53 ps  
90804A  
54 ps  
38 ps  
91204A  
35 ps  
25 ps  
91304A  
32 ps  
23 ps  
Input coupling  
Vertical resolution 3  
8 bits, ≥ 12 bits with averaging  
Channel to channel isolation (any two  
channels with equal V/div settings  
DC to 3 GHz: 90804A/91204A/91304A: 60 dB (≥ 1000:1)  
90254A/90404A/90604A: 50 dB (≥ 316:1)  
3 to 8 GHz: 40 dB (≥ 100:1)  
8 GHz to BW: 35 dB (≥ 56:1)  
± 2% of full scale at full resolution channel scale (± 2.5% for 5mV/div)  
± 5 V  
DC gain accuracy 1, 2  
Maximum input voltage 1  
Offset range  
Vertical sensitivity  
Available offset  
± 0.4 V  
± 0.9 V  
± 1.6 V  
± 3.0 V  
0 mV/div to ≥ 40 mV/div  
> 40 mV/div to ≥ 75 mV/div  
> 75 mV/div to ≥ 130 mV/div  
> 130 mV/div to ≥ 240 mV/div  
> 240 mV/div  
≤ 3.5 V: ± (2% of channel offset + 1% of full scale) + 1 mV  
> 3.5 V: ± (2% of channel offset + 1% of full scale)  
± 4 div from center screen  
± 4.0 V  
Offset accuracy 1, 2  
Dynamic range  
DC voltage measurement accuracy 1, 2  
Dual cursor: ± [(DC gain accuracy) + (resolution)]  
Single cursor: ± [(DC gain accuracy) + (offset accuracy) + (resolution/2)]  
RMS noise floor (scope only)  
Volts/div  
5 mV  
90254A  
153 μV  
90404A  
199 μV  
90604A  
259 μV  
295 μV  
424 μV  
985 μV  
1.92 mV  
3.71 mV  
9.91 mV  
19.2 mV  
90804A  
322 μV  
358 μV  
498 μV  
1.15 mV  
2.22 mV  
4.28 mV  
11.5 mV  
22.3 mV  
91204A  
435 μV  
483 μV  
650 μV  
1.45 mV  
2.80 mV  
5.41 mV  
14.7 mV  
28.5 mV  
91304A  
467 μV  
536 μV  
758 μV  
1.73 mV  
3.37 mV  
6.58 mV  
17.4 mV  
34.1 mV  
10 mV  
20 mV  
50 mV  
100 mV  
200 mV  
500 mV  
1 V  
183 μV  
275 μV  
645 μV  
1.27 mV  
2.47 mV  
6.48 mV  
12.5 mV  
232 μV  
342 μV  
799 μV  
1.56 mV  
3.03 mV  
8.00 mV  
15.6 mV  
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period, and ± 5 °C from annual calibration  
temperature�  
2. Full scale is defined as 8 vertical divisions. Magnification is used below 5 mV/div. Below 5 mV/div, full-scale is defined as 40 mV. The major scale settings are  
5 mV, 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV, 1 V�  
3� Vertical resolution for 8 bits = 0�4% of full scale, for 12 bits = 0�024% of full scale�  
4� 13 GHz DSP enhanced bandwidth not applicable at 5 mV/div�  
5. FFT amplitude readings are affected by scope and probe bandwidth limitations and input amplifiers roll-off (e.g. 3 dB roll-off at specified bandwidth of  
scope/probe)�  
6� The FFT signal to noise ratio varies with volts/division setting, memory depth and use of time or frequency averaging�  
7a. Noise is the displayed noise floor. SlewRate is the displayed slew rate of the signal at the threshold crossings. Sample rate = max, sin(x)/x interpolation  
enabled�  
7b. Measurement threshold = fixed voltage at 50% level.  
7c. Time ranges ≤ 10 μs.  
8� Values represent time error between two edges on a single channel� Standard deviation value refers to the standard deviation of 256 consecutive  
measurements performed using an individual instrument. Reading is the displayed DTMA measurement value. TimeScaleAccy is the oscilloscope’s specified  
time scale accuracy�  
9. Internal edge trigger mode. Trigger threshold = fixed voltage at 50% level. The slew rate independent value in the formula represents the traditional trigger  
jitter�  
10. Requires Option 010 user defined function.  
11� 11�8 GHz analog bandwidth at 5 mV/div for DSO91304A and DSO91204A models�  
12� Typically triggers as low as 5 mV/div sensitivity�  