12 | Keysight | InfiniiVision 6000L Series Low-Profile Oscilloscopes - Data Sheet
Performance characteristics (continued)
General characteristics (continued)
Trigger out
When Triggers is selected
(delay ~17 ns)
0 to 5 V into high impedance
0 to 2.5 V into 50 Ω
When Source Frequency or
Source Frequency/8 is selected
0 to 580 mV into high impedance
0 to 290 mV into 50 Ω
Max frequency output
10 MHz ref in/out
350 MHz (in source frequency mode when terminated in 50 Ω)
125 MHz (in source frequency/8 mode when terminated in 50 Ω)
TTL out, 180 mV to 1 V amplitude within 0 to 2 V offset
Power requirements
Line voltage range
100-240 V, 50/60 Hz auto ranging
Line frequency
Power usage
50/60 Hz
80 W max
Environmental characteristics
Ambient temperature
Operating -10 ꢁC to +50 ꢁC; non-operating -40 ꢁC to +70 ꢁC
Humidity
Altitude
Vibration
Operating 95% RH at 40 ꢁC for 24 hours; Non-operating 90% RH at 65 ꢁC for 24 hours
Operating to 4,570 m (15,000 ft); non-operating to 15,244 m (50,000 ft)
Keysight class GP and MIL-PRF-28800F; Class 3 random
Shock
axis
Keysight class GP and MIL-PRF-28800F; (operating 30 g, 1/2 sine, 11-ms duration, 3 shocks/
along major axis. Total of 18 shocks)
Pollution degree
Indoor use
Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation must be expected.
This instrument is rated for indoor use only
Other
Installation categories
CAT I
EMC
IEC 61326-1:1997, EN 61326-1:1997
Safety
IEC 61010-1:2001, EN 61010-1:2001
Canada: CSA-C22.2 No. 1010.1:1992
UL 61010-1:2003
Supplementary information
TheproductherewithcomplieswiththerequirementsoftheLowVoltageDirective73/23/EEC