42 | Keysight | B1505A Power Device Analyzer/Curve Tracer - Data Sheet  
Physical constants  
Keyboard constants are stored in memory  
as follows:  
q: Electron charge, 1.602177E-19 C  
k: Boltzmann’s constant, 1.380658E-23  
ε (e): Dielectric constant of vacuum,  
8.854188E-12  
Data Protection  
About measurement accuracy  
EasyEXPERT group+ has various options to RF electromagnetic field and SMU  
protect important data as follows.  
– Password protection (workspace, test  
definition and my favorite)  
– User level access control (engineer  
mode/operator mode)  
measurement accuracy: SMU volt-  
age and current measurement accuracy  
can be affected by RF electromagnetic  
field strengths greater than 3 V/m in the  
frequency range of 80 MHz to 1 GHz. The  
extent of this effect depends upon how the  
instrument is positioned and shielded.  
Engineering units  
Workspace back-up and porta-  
bility  
EasyEXPERT group+ has the ability to  
import/export a workspace for back-up and  
portability.  
The following unit symbols are also avail-  
able on the keyboard:  
a (10-18), f (10-15), p (10-12), n (10-9),  
u or μ (10-6), m (10-3), k (103),  
M (106), G (109), T (1012) , P (1015)  
Induced RF field noise and SMU  
measurement accuracy:  
SMU voltage and current measurement  
accuracy can be affected by  
induced RF field noise strengths  
greater than 3 Vrms in the frequency  
range of 150 kHz to 80 MHz. The  
extent of this effect depends upon  
how the instrument is positioned  
and shielded.  
Data management  
General specification  
Workspace (Built-in database)  
EasyEXPERT group+ supports the built-in  
database called “workspace”. Workspaces  
are created on a HDD, and they enable to  
manage and access all the measurement  
related data without handling numerous  
files. Every workspace supports the fol-  
lowing features:  
– Access to measurement capabilities  
and data stored in the workspace.  
Save/Import/Export measurement  
settings and data (application library,  
measurement settings, my favorite  
setup, and measurement data)  
Temperature range  
Operating: +5 °C to +40 °C  
Storage: -20 °C to +60 °C  
Regulatory compliance  
EMC:  
IEC 61326-1 / EN 61326-1  
Canada: ICES/NMB-001  
AS/NZS CISPR 11  
Humidity range1  
Operating: 20% to 70% RH,  
non-condensing  
Storage: 10% to 90% RH,  
non-condensing  
Safety:  
IEC61010-1 / EN 61010-1  
CAN/CSA-C22.2 No. 61010-1  
Storage: 20% to 80% RH,  
non-condensing (N1268A)  
Certification  
CE, cCSAus, RCM  
Recall the setup for measurement  
reproduction and data for analysis  
Altitude  
Operating: 0 m to 2,000 m (6,561 ft)  
Storage: 0 m to 4,600 m (15,092 ft)  
0 m to 2,000 m (6,561 ft) (N1268A)  
Dimensions  
B1505A:  
Data auto record/ auto export  
EasyEXPERT group+ has the ability to au-  
tomatically store the measurement setup  
and data within a workspace. It can also  
export measurement data in real time, in a  
variety of formats such as Excel (xls).  
420 mm W x 330 mm H x 575 mm D  
N1258A module selector:  
330 mm W x 120 mm H x 410 mm D  
N1259A test fixture:  
420 mm W x 272 mm H x 410 mm D  
N1260A High voltage bias-tee:  
164 mm W x 53 mm H x 125 mm D  
N1261A-001 HPSMU protection adapter  
(Triaxial output):  
80 mm W x 40 mm H x 110 mm D  
N1261A-002 GNDU protection adapter  
(BNC output):  
Power requirement  
ac Voltage: 90 V to 264 V  
Line Frequency: 47 Hz to 63 Hz  
Maximum volt-amps (VA)  
B1505A: 900 VA  
N1258A: 65 VA  
N1259A-300: 35 VA  
N1265A: 400 VA  
N1266A: 60 VA  
Import/export files  
File type:  
Keysight EasyEXPERT format, XML-SS  
format, CSV format  
N1268A: 350 VA  
N1272A: 70 VA  
80 mm W x 40 mm H x 110 mm D  
N1261A-003 HPSMU protection adapter  
(HV triaxial output):  
90 mm W x 40 mm H x 140 mm D  
N1261A-004 GNDU protection adapter  
(SHV output):  
Acoustic Noise Emission  
Lpa < 65 dB  
Lwa: 66 dB (Operating mode)  
Lwa: 73 dB (Worst case mode)  
80 mm W x 40 mm H x 125 mm D  