23 | Keysight | Infiniium DCA-X 86100D Wide-Bandwidth Oscilloscope Mainframe and Modules - Data Sheet  
Modules Specifications (Continued)  
Clock recovery  
The DCA platform offers clock recovery capability in the form of DCA plug-in modules such as the 83496B or standalone clock  
recovery solutions such as the N1076A/N1077A. Specifications for 83496B modules are summarized below. For more information on  
standalone clock recovery solutions, please download the N1076A/N1077A data sheet from www.keysight.com/find/N1077A.  
Clock recovery modules  
83496B-100  
83496B-101  
Channel type  
Differential or single-ended electrical  
Single-mode or multimode optical, differential or single-ended  
electrical (no internal electrical splitters)  
Data rates (divide by 2 for clock signals)  
Standard: 50 Mb/s to 7.1 Gb/s continuous tuning  
Option 200: 50 Mb/s to 14.2 Gb/s continuous tuning  
Option 201: 7.1 to 14.2 Gb/s continuous tuning  
For rates above 14 Gb/s, the N1076A and N1077A operate from 50 Mb/s to 32 Gb/s  
Minimum input level to acquire lock  
(voltage or OMA 1)  
150 m Vpp  
Single-mode (OMA 1):  
–11 dBm to 50 Mb/s to 11.4 Gb/s  
–8 dBm to > 11.4 Gb/s  
–12 dBm to 7.1 to 14.2 Gb/s (with Option 200)  
–14 dBm to 1 to 7.1 Gb/s  
–15 dBm to 50 Mb/s to 1 Gb/s  
Multimode 1310 nm (OMA 1):  
–10 dBm to 50 Mb/s to 11.4 Gb/s  
–7 dBm to > 11.4 Gb/s  
–11 dBm to 7.1 to 14.2 Gb/s (with Option 200)  
–13 dBm to 1 to 7.1 Gb/s  
–14 dBm to 50 Mb/s to 1 Gb/s  
Multimode 850 nm (OMA 1):  
–8 dBm to 50 Mb/s to 11.4 Gb/s  
–7 dBm to > 11.4 Gb/s  
–9 dBm to 7.1 to 14.2 Gb/s (with Option 200)  
–11 dBm to 1 to 7.1 Gb/s  
–12 dBm to 50 Mb/s to 1 Gb/s  
Electrical: 150 m Vpp  
Output random jitter (RMS) 2  
Internal recovered clock trigger:  
< 500 fs 7.2 to 11.4 Gb/s (300 fs to 10 Gb/s)  
< 700 fs 4.2 to 7.2 Gb/s, 11.4 Gb/s to 14.2 Gb/s (400 fs to 4.25 Gb/s, 500 fs to 2.5 Gb/s)  
< 3 mUI 50 Mb/s to 4.2 Gb/s (700 fs to 1.25 Gb/s)  
Front panel recovered clock:  
< 700 fs 7.2 to 11.4 Gb/s (300 fs to 10 Gb/s)  
< 900 fs 4.2 to 7.2 Gb/s, 11.4 to 14.2 Gb/s (400 fs to 4.25 Gb/s, 500 fs to 2.5 Gb/s)  
< 4 mUI 50 Mb/s to 4.2 Gb/s (700 fs to 1.25 Gb/s)  
Clock recovery adjustable loop  
bandwidth range (user selectable)  
Loop bandwidth accuracy  
Standard: 270 kHz or 1.5 MHz 3  
Option 300: 15 kHz to 10 MHz 4 continuous tuning (fixed value or a constant rate/N ratio)  
Standard: ± 30%  
Option 300: ± 25% for transition density = 0.5 and data rate 155 Mb/s to 11.4 Gb/s  
(± 30% for 0.25 ≤ transition density ≤ 1.0 and all data rates)  
Tracking range  
Aquisition range  
Internal splitter ratio  
± 2500 ppm, 83496B  
± 5000 ppm  
50/50  
± 1000 ppm, 83496A  
50/50 single-mode  
30/70 multimode  
Electrical signals have input only (no internal power dividers)  
20 dB single-mode, 16 dB multimode  
22 dB min (DC to 12 GHz) electrical  
16 dB min (12 to 20 GHz) electrical  
4 dB max single-mode optical  
Input return loss  
22 dB (DC to 12 GHz) electrical  
16 dB (12 to 20 GHz) electrical  
Input insertion loss  
7.2 dB max (DC to 12 GHz) electrical  
7.8 dB max (12 to 20 GHz) electrical  
4 dB max multimode optical (no electrical data output signal path)  
See footnotes on next page.  