03 | Keysight | 6610C Series Single-Output, 40-50 W GPIB Power Supplies - Data Sheet
Speciꢀcations
Specifications
(at 0 ° to 55 ° unless otherwise specified)
6611C-J05
Special Order
Option
6611C
6612C
6613C
6614C
Number of outputs
GPIB
1
1
1
1
1
Yes
Yes
Yes
Yes
Yes
Output ratings
Voltage
Current
0 to 8 V
0 to 5 A
0 to 20 V
0 to 2 A
0 to 50 V
0 to 1 A
0 to 100 V
0 to 0.5 A
0 to 10 V
0 to 5 A
Programming accuracy at 25 °C 5 °C
Voltage
Current
5 mV
2 mA
10 mV
1 mA
20 mV
0.75 mA
50 mV
0.5 mA
5 mV
2 mA
0.05% +
Ripple and noise
20 Hz to 20 MHz;with outputs ungrounded
or with either terminal grounded
Voltage
rms
0.5 mV
3 mV
2 mA
0.5 mV
3 mV
1 mA
0.5 mV
4 mV
1 mA
0.5 mV
5 mV
1 mA
0.5 mV
3 mV
2 mA
peak-to-peak
rms
Normal mode
DC measurement accuracy
via GPIB or front-panel meters with respect
to actual output at 25 °C 5 °C
Voltage
0.03% +
2 mV
3 mV
6 mV
1
2 mV
2 mV
Low current range
–20 mA to + 20 mA
High current range
+20 mA to + rated I
–20 mA to – rated I
0.1% +
2.5 µA
2.5 µA
2.5 µA
2.5 µA
2.5 µA
0.2% +
0.2% +
0.5 mA
1.1 mA
0.25 mA
0.85 mA
0.2 mA
0.8 mA
0.1 mA
0.7 mA
0.5 mA
1.1 mA
Load regulation
Voltage
Current
2 mV
1 mA
2.5 mV
0.5 mA
1 mV
2 mA
2.5 mV
1 mA
2 mV
1 mA
Line regulation
Voltage
Current
0.5 mV
0.5 mA
0.5 mV
0.5 mA
1 mV
0.25 mA
1 mV
0.25 mA
0.5 mV
0.5 mA
Transient response time
Less than 100 µs for the output to recover to its previous level (within 0.1% of the voltage rating of the supply or 20 mV, whichever is greater) following any step
change in load current of up to 50% of the output current rating of the supply
Supplemental Characteristics
(Non-warranted characteristics determined by design
and useful in applying the product)
6611C-J05
Special order
option
6611C
6612C
6613C
6614C
Average programming resolution
Voltage
Current
2 mV
1.25 mA
5 mV
0.5 mA
12.5 mV
0.25 mA
25 mV
0.125 mA
3 mV
1.25 mA
Sink current
3 A
1.2 A
0.6 A
0.3 A
3 A