03 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet  
Technical Specifications  
Get more information  
Definitions and conditions  
Temperatures referred to in this document are defined as follows:  
– Full temperature range = Individual module temperature of 5 to 68 °C, as reported by  
the module, and environment temperature of 0 to 55 °C.  
– Controlled temperature range = Individual module temperature of 25 to 40 °C, as  
reported by the module, and environment temperature of 20 to 30 °C.  
This CXA-m PXIe X-Series signal  
analyzer data sheet is a summary  
of the complete specifications  
and conditions available in the  
CXA-m PXIe Signal Analyzer  
Specification Guide. The CXA-m  
PXIe Signal Analyzer Specifica-  
tion Guide can be obtained on  
the web at:  
Specifications describe the warranted performance of calibrated instruments. Data  
represented in this document are specifications under the following conditions unless  
otherwise noted.  
– It is within its calibration cycle  
www.keysight.com/find/  
cxa-m_manuals  
– Under auto couple control, except when Auto Sweep Time Rules = Accy  
– The analyzer has been stored at an ambient temperature within the allowed  
operating range for at least two hours before being turned on; if it had previously  
been stored at a temperature range inside the allowed storage range, but outside  
the allowed operating range  
For ordering information, refer to  
the CXA-m PXIe Signal Analyzer  
Configuration Guide  
– The analyzer has been turned on at least 30 minutes with Auto Align set to normal,  
or, if Auto Align is set to off or partial, alignments must have been run recently  
enough to prevent an Alert message; if the Alert condition is changed from Time and  
Temperature to one of the disabled duration choices, the analyzer may fail to meet  
specifications without informing the user  
(5992-0193EN).  
95th percentile values indicate the breadth of the population (approx. 2σ) of perfor-  
mance tolerances expected to be met in 95 percent of the cases with a 95 percent  
confidence, for any ambient temperature in the range of 20 to 30 °C. In addition to the  
statistical observations of a sample of instruments, these values include the effects of  
the uncertainties of external calibration references. These values are not warranted.  
These values are updated occasionally if a significant change in the statistically observed  
behavior of production instruments is observed.  
Typical describes additional product performance information that is not covered by  
the product warranty. It is performance beyond specifications that 80 percent of the  
units exhibit with a 95 percent confidence level over the temperature range 20 to 30 °C.  
Typical performance does not include measurement uncertainty.  
Nominal values indicate expected performance, or describe product performance that is  
useful in the application of the product, but are not covered by the product warranty.  
Recommended best practices in use  
– Use slot blockers and EMC filler panels in empty module slots to ensure proper  
operating temperatures. Keysight chassis and slot blockers optimize module  
temperature performance and reliability of test.  
– Set chassis fan to high at environmental temperatures above 45°C.