14 | Keysight | M9290A CXA-m PXIe X-Series Signal Analyzer - Data Sheet
General Specifications
Temperature range
Operating
Storage
EMC
0 to 55 °C
–40 to 70 °C
Complies with European EMC Directive 2004/108/EC
– IEC/EN 61326-1
– CISPR Pub 11 Group 1, class A
– AS/NZS CISPR 11
– ICES/NMB-001
This ISM device complies with Canadian ICES-001
Cet appareil ISM est conforme à la norme NMB-001 du Canada
Environmental stress
Samples of this product have been type tested in accordance with the Keysight Environmental Test Manual and verified to be robust against the environ-
mental stresses of storage, transportation, and end-use; those stresses include, but are not limited to, temperature, humidity, shock, vibration, altitude,
and power line conditions; test methods are aligned with IEC 60068-2 and levels are similar to MILPRF-28800F Class 3.
Power requirement
Power drawn from chassis
Weight
≤ 65 W
Net
Shipping
1.9 kg (4.2 lbs)
4.2 kg (9.3 lbs)
Dimensions
Height
Width
132 mm (5.2 in)
82 mm (3.2 in)
202 mm (8.0 in)
Length
Calibration cycle
The recommended calibration cycle is one year; calibration services are available through Keysight service centers
Inputs and Outputs
RF input
Connector
RF output
Connector
10 MHz in
Connector
10 MHz out
Connector
Trigger in
Connector
Trigger out
Connector
Analog out
Connector
Noise source
3.5mm female, 50 Ω nominal
3.5mm female, 50 Ω nominal
SMB male, 50 Ω nominal
SMB male, 50 Ω nominal
SMB male, 10 kΩ nominal
SMB male, 50 Ω nominal
SMB male, 50 Ω nominal
For use with Keysight 346 and SNS series noise sources