20 | Keysight | 4285A Precision LCR Meter – Data Sheet
Display time
Display time for each display format is given as:
MEAS DISPLAY page
BIN No. DISPLAY page
Approx. 8 ms
Approx. 5 ms
BIN COUNT DISPLAY page Approx. 0.5 ms
GPIB data output time
Internal GPIB data, processing time from EOM output to measurement data output on
GPIB lines (excluding display time):
–
Approx. 10 ms
Option 4285A-001 (internal DC bias)
Maximum DC bias current when the normal measurement can be performed is 100 mA.
DC bias settling time
When DC bias is set to ON, add 5 ms to the measurement time. This settling time does
not include the DUT charge time.
Sum of DC bias settling time plus DUT (capacitor) charge time is shown in the following
figure.
Figure 1-6. Sum of the DC bias settling time and DUT (capacitor) charge time