14 | Keysight | 34980A Multifunction Switch/Measure Unit - Data Sheet  
Table 4. Matrix selection table—speciꢁcations and characteristics  
34931A  
34932A  
34933A  
34934A  
Channels/conꢀgurations  
dual 4x8  
8x8  
dual 4x16  
8x16  
dual 4x8  
8x8  
quad 4x32  
4x128  
4x16  
4x32  
4x16  
8x64  
quad 4x8, 1-wire  
16x32  
Switch type  
Armature  
latching  
Armature  
latching  
Reed  
non-latching  
Reed  
non-latching  
Input characteristics (per channel)  
Max volts  
± 300 V [1]  
± 300 V [1]  
± 150 V peak [2]  
± 100 V peak  
Max current (DC, AC RMS)  
Switch current  
Carry current  
1 A  
2 A  
1 A  
2 A  
0.5 A [5]/0.05 A [8]  
1.5 A [5]/0.05 A [8]  
0.5 A  
0.5 A  
Power (W, VA) [2, 6]  
60 W  
108  
60 W  
108  
10 W [7]  
108  
10 W  
108  
Volt-Hertz limit  
Initial closed channel res [3][9]  
< 1.5 Ω  
< 1.5 Ω  
<
1.5 Ω [5]/200 Ω [8]  
< 1Ω/100 Ω  
nominal  
General Speciꢀcations  
Offset voltage [3]  
< 3 uV  
< 3 uV  
< 50 uV  
< 20 uV  
< 100 uV 1-wire  
< 50 uV 1-wire  
DC Isolation (ch-ch, ch-earth)  
AC characteristics  
> 10G Ω  
> 10G Ω  
> 10G Ω  
10G Ω  
Bandwidth at terminal block [4]  
30 MHz  
30 MHz  
30 MHz [5]/4 MHz [8] 35 MHz 2-wire  
2 MHz 1-wire  
15 MHz 1-wire  
Crosstalk at terminal block (ch-ch) [4]  
300 kHz  
1 MHz  
20 MHz  
–65 dB  
–55 dB  
–30 dB  
–65 dB  
–55 dB  
–30 dB  
–65 dB  
–55 dB  
–40 dB  
–65 dB  
–55 dB  
–33 dB  
Capacitance at terminal block  
HI-LO  
LO – earth  
50 pF  
80 pF  
50 pF  
80 pF  
80 pF  
75 pF  
45 pF  
250 pF  
General characteristics  
Relay life, typical  
No load  
10 V, 100 mA  
Rated load  
100 M  
10 M  
100 k  
100 M  
10 M  
100 k  
1000 M  
10 M  
10 k  
1000 M operations  
Open/close time  
4 ms/4 ms  
Bank 2  
4 ms/4 ms  
Bank 2  
0.5 ms/0.5 ms  
Bank 2  
0.35 ms/0.10 ms  
No  
Analog bus backplane connection  
[1] DC or AC RMS voltage, channel-to-channel or channel-to-earth  
[2] Peak voltage, channel-to-channel or channel-to-earth  
[3] Into analog bus. System errors are included in the internal DMM measurement accuracy specifications  
[4] 50 Ω source, 50 Ω load, differential measurements verified (Sdd21)  
[5] With input resistors bypassed. Bypassing resistors will reduce lifetime of relays. See the rated load relay life characteristics.  
[6] Limited to 6 W channel resistance power loss per module  
[7] Power restrictions allow only 20 channels to be closed at one time  
[8] Protection Resistors:  
34933A - 100Ω ±5%; 0.5W; TC = ±200ppm/°C.  
34934A - 100Ω ±1%; 0.25W; TC = ±100ppm/°C.  
If this resistance is not bypassed in the low side source line of a 4-wire resistance measurement, the 100 Ω range is limited.  
[9] Channel resistance is typically < 1.5 Ω but can go as high as 50 Ω when a channel is used in measurement applications with < 10 mA  
load current. Increased relay channel resistance for measurements with load currents below 10 mA can occur on cards that have been  
out of service or following relay inactivity for periods of greater than 1 week. Switching relays for 2K cycles prior to use may reduce  
the variation in channel resistance. Applies to the 34931A and 34932A. Keysight recommends the use of 4-wire Ohms for resistance  
measurements. For high accuracy voltage measurements, select the DMM input resistance setting of > 10 G ohms to minimize the  
impact of relay contact resistance.  