1
Test port output power  
Table 12. Option 230/235/240/245/260/265/280/285/430/435/440/  
445/460/465/480/485  
Description  
Specification  
Typical  
Nominal power  
(preset power)  
0 dBm  
Level accuracy2, 6  
(stepped sweep mode)  
0.650 dB (at 0 dBm,  
50 MHz absolute)  
1.0 dB  
Level accuracy2  
(swept sweep mode)  
2.5 dB  
Level linearity3, 5, 6  
(stepped sweep mode)  
9 kHz to 5 GHz  
0.75 dB (–20 to 10 dBm)  
0.75 dB (–20 to 9 dBm )  
0.75 dB (–20 to 8 dBm )  
0.75 dB (–20 to 7 dBm )  
5 GHz to 6 GHz  
6 GHz to 7 GHz  
7 GHz to 8.5 GHz  
Level linearity5  
(swept sweep mode)4  
9 kHz to 5 GHz  
1.5 dB (at –20 to 10 dBm)  
1.5 dB (at –20 to 9 dBm)  
1.5 dB (at –20 to 8 dBm)  
1.5 dB (at –20 to 7 dBm)  
5 GHz to 6 GHz  
6 GHz to 7 GHz  
7 GHz to 8.5 GHz  
Range5, 6  
–55 to 10 dBm  
–55 to 9 dBm  
–55 to 8 dBm  
–55 to 7 dBm  
9 kHz to 5 GHz  
5 GHz to 6 GHz  
6 GHz to 7 GHz  
7 GHz to 8.5 GHz  
Sweep range5, 6  
9 kHz to 5 GHz  
5 GHz to 6 GHz  
6 GHz to 7 GHz  
7 GHz to 8.5 GHz  
–55 to 10 dBm  
–55 to 9 dBm  
–55 to 8 dBm  
–55 to 7 dBm  
Level resolution  
0.05 dB  
Description  
Specification  
SPD  
Max leveled power5, 6  
(Option 230/235/240/245/260/  
265/280/285)  
9 kHz to 8.5 GHz  
10 dBm  
(Option 430/435/440/445/460/  
465/480/485)  
9 kHz to 6 GHz  
10 dBm  
9 dBm  
8 dBm  
6 GHz to 7 GHz  
7 GHz to 8.5 GHz  
1. Source output performance on port 1 only. Other port output performance is typical.  
2. Level accuracy is taken at 0 dBm, relative to 50 MHz reference unless otherwise stated.  
3. Level linearity given is relative to 0 dBm.  
4. The specification might not be met at 5 MHz or 50 MHz.  
5. The level accuracy specification needs to be taken into account for test port output power level.  
6. Power calibration using an external power meter improves level accuracy of the test port output power.  
Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration.  