Corrected System Performance
The specifications in this section apply to measurements made with the
Agilent E5071C network analyzer under the following conditions:
•
•
No averaging applied to data
Environmental temperature of 23 °C ( 5 °C) with less than 1 °C deviation
from the calibration temperature
•
Response and isolation calibration performed
System dynamic range
Table 1. Option 230/235/240/245/260/265/280/285/430/435/440/445/
460/465/480/485
Description
Specification
SPD
System dynamic range1, 2
9 kHz to 300 kHz
72 dB
300 kHz to 10 MHz
82 dB
IF bandwidth = 3 kHz
10 MHz to 6 GHz
6 GHz to 8.5 GHz
9 kHz to 300 kHz
300 kHz to 10 MHz
10 MHz to 6 GHz
6 GHz to 7 GHz
98 dB
92 dB
97 dB
115 dB
115 dB
130 dB
128 dB
126 dB
124 dB
107 dB
123 dB
117 dB
117 dB
117 dB
IF bandwidth = 10 Hz
7 GHz to 8 GHz
8 GHz to 8.5 GHz
160
150
140
130
120
110
100
90
E5071C specification
80
1.E+05
1.E+09
2.E+09
3.E+09
4.E+09
5.E+09
6.E+09
7.E+09
8.E+09
Test frequency [Hz]
Figure 1. System dynamic range (specification and actual measurement data example,
IF bandwidth 10 Hz)
1. The test port dynamic range is calculated as the difference between the test port rms noise floor
and the source maximum output power. The effective dynamic range must take measurement
uncertainty and interfering signals into account.
2. The specification might not be met at 5 MHz or 50 MHz.
5