Test port output power (continued)
Table 13. Option 2D5/2K5/4D5/4K5
Description
Specification
SPD
Max leveled power3, 4
300 kHz to 1 MHz
1 MHz to 10 GHz
10 GHz to 13 GHz
13 GHz to 15 GHz
15 GHz to 18 GHz
18 GHz to 20 GHz
9 dBm
10 dBm
9 dBm
7 dBm
5 dBm
4 dBm
Test port output signal purity
Table 14. Option 230/235/240/245/260/265/280/285/430/435/440/
445/460/465/480/485
Description
Specification
Typical
Harmonics (2nd or 3rd)
9 kHz to 2 GHz
< –25 dBc (at 5 dBm)
< –20 dBc (at 5 dBm)
2 GHz to 8.5 GHz
Non-harmonic spurious
9 kHz to 8.5 GHz
< –30 dBc (at 5 dBm)
Table 15. Option 2D5/2K5/4D5/4K5
Description
Specification
Typical
Harmonics (2nd to 5th)1
300 kHz to 1 GHz
< –12 dBc (at maximum
output power)5
1 GHz to 20 GHz
< –15 dBc (at maximum
output power)5
Non-harmonic spurious2
300 kHz to 20 GHz
< –30 dBc (at –5 dBm)
1. This includes 6th and 7th harmonics when the test frequency is from 1 MHz to 1 GHz.
1
2. The carrier
/8th spurious is excluded from 8.76 GHz to 17.52 GHz.
3. The level accuracy specification needs to be taken into account for test port output power level.
4. Power calibration using an external power meter improves level accuracy of the test port output power.
Proper power meters/sensors, and the 82357B USB-GPIB interface are required to conduct power calibration.
5. Maximum output power is the maximum power of "Range" in Table 13 Test port output power.
6. Source output performance on port 1 only. Other port output performance is typical.