Frequency Specifications  
Frequency range  
E4443A  
E4445A  
E4440A  
E4447A  
E4446A  
E4448A  
(DC coupled)  
(DC coupled)  
(DC coupled)  
(DC coupled)  
(DC coupled)  
(DC coupled)  
3 Hz to 6.7 GHz  
3 Hz to 13.2 GHz  
3 Hz to 26.5 GHz1  
3 Hz to 42.98 GHz  
3 Hz to 44 GHz1  
3 Hz to 50 GHz1  
(AC coupled)  
(AC coupled)  
(AC coupled)  
20 MHz to 6.7 GHz  
20 MHz to 13.2 GHz  
20 MHz to 26.5 GHz1  
1. 325 GHz with external mixers  
Band  
Harmonic mixing mode (N)  
0
1
2
3
4
5
6
1–  
1–  
2–  
4–  
4–  
4+  
8–  
3 Hz to 3 GHz  
2.85 GHz to 6.6 GHz  
6.2 GHz to 13.2 GHz  
12.8 GHz to 19.2 GHz  
18.7 GHz to 26.8 GHz  
26.4 GHz to 31.15 GHz  
31.0 GHz to 50.0 GHz  
Frequency reference  
Accuracy  
± ±(time since last adjustment x aging rate) + temperature stability + calibration accuracyꢀ  
± 1 x 10–7 / year  
Aging rate  
Temperature stability  
20 °C to 30 °C ± 1 x 10–8  
± 7 x 10–8  
0 °C to 55 °C ± 5 x 10–8  
Achievable initial calibration accuracy  
Example frequency reference accuracy  
1 year after last adjustment  
= ± (1 x 1 x 10–7 + 1 x 10–8 + 7 x 10–8)  
= ± 1.8 x 10–7  
Frequency readout accuracy (start, stop, center, marker)  
± (marker frequency x frequency reference accuracy + 0.25ꢁ x span + 5ꢁ x RBW + 2 Hz + 0.5 x horizontal resolution*)  
* Horizontal resolution is span/(sweep points – 1)  
Marker frequency counter  
Accuracy  
± (marker frequency x frequency reference accuracy + 0.100 Hz)  
± (delta frequency x frequency reference accuracy + 0.141 Hz)  
0.001 Hz  
Delta counter accuracy  
Counter resolution  
Frequency span (FFT and swept mode)  
Range  
0 Hz (zero span), 10 Hz to maximum frequency of model  
2 Hz  
Resolution  
Accuracy  
± ±0.2ꢁ x span + span / (sweep points – 1)ꢀ  
4