FEATURES
ɀ Measures positive and negative gage ɀ Medical grade ISO10993-1(USP Class
pressures
6) port material
ɀ Flow-through port design fits in-line ɀ Silicon sensor chip
with application
ɀ Popular port sizes:
- 8 mm (.315 in.) OD (1/4 in. ID tubing or ɀ Minimal deadspace — efficient cleans-
standard connectors)
- 0.144 in. OD (1/8 in. ID tubing)
ɀ 24 inch wire harness with splash proof
connector
ing and disinfecting
24PC SERIES PERFORMANCE CHARACTERISTICS at 10.0 ±0.01 VDC
EXCITATION, 25°C
Min.
---
Typ.
10
Max.
12
Units
VDC
Excitation
Null Shift, 25° to 0°, 25° to 50°C
Null Offset
---
±2.0
0
---
mV
−30
---
+30
---
mV
Linearity, P2 > P1, BFSL
Span Shift, 25° to 0°, 25° to 50°C
Repeatability & Hysteresis
Response Time
±0.5
±5.0
±0.2
---
%Span
%Span
%Span
msec
ohms
ohms
%Span
---
---
---
---
---
1.0
6.0 K
6.0 K
---
Input Resistance
4.0 K
4.0 K
---
5.0 K
5.0 K
±0.5
Output Resistance
Stability over One Year
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Shock
−40° to +85°C (−40° to +185°F)
−55° to +100°C (−67° to +212°F)
Qualification tested to 150 g
Vibration
Qualification tested to 0 to 2 kHz, 20 g sine
Media Compatibility
Limited only to those media which will not attack
polysulfone, silicon, fluorosilicone, silicone,
EPDM, and neoprene seals
24PC SERIES FLOW THROUGH ORDER GUIDE
Pressure
Sensitivity
mV/psi
Typ.
Span, mV
Typ.
35
Catalog
Listing
Range
psi
Overpressure
psi Max.
Min.
25
Max.
45
24PCE Type
24PCA Type
24PCB Type
24PCC Type
24PCD Type
24PCF Type
24PCG Type
0.5
1.0
5.0
15
70
45
20
20
30
45
60
85
115
145
285
420
294
280
23
20
165
240
156
145
225
15
45
30
330
11
60
100
250
225
2.25
0.85
200
500
212
Honeywell ɀ Sensing and Control ɀ1-800-537-6945 USA ɀ ⍣1-815-235-6847 International ɀ1-800-737-3360 Canada