Pressure Sensors
Gage Unamplified Noncompensated Flow-Through
FEATURES
Measures positive and negative gage
pressures
Flow-through port design fits in-line
with application
Popular port sizes:
- 8 mm (.315 in.) OD (1/4 in. ID tubing or
standard connectors)
- 0.144 in. OD (1/8 in. ID tubing)
24PC SERIES PERFORMANCE CHARACTERISTICS at 10.0
±0.01
VDC
EXCITATION, 25°C
Min.
Excitation
Null Shift, 25° to 0°, 25° to 50°C
Null Offset
Linearity, P2
>
P1, BFSL
Span Shift, 25° to 0°, 25° to 50°C
Repeatability & Hysteresis
Response Time
Input Resistance
Output Resistance
Stability over One Year
---
---
−30
---
---
---
---
4.0 K
4.0 K
---
Typ.
10
±2.0
0
±0.5
±5.0
±0.2
---
5.0 K
5.0 K
±0.5
Max.
12
---
+30
---
---
---
1.0
6.0 K
6.0 K
---
Units
VDC
mV
mV
% Span
% Span
% Span
msec
ohms
ohms
% Span
24PC Series
Medical grade ISO 10993-1 (USP Class
6) port material
Silicon sensor chip
24 inch wire harness with splash proof
connector
Minimal deadspace — efficient cleans-
ing and disinfecting
Unamplified
ENVIRONMENTAL SPECIFICATIONS
Operating Temperature
Storage Temperature
Shock
Vibration
Media Compatibility
−40°
to
+85°C
(−40° to
+185°F)
−55°
to
+100°C
(−67° to
+212°F)
Qualification tested to 150 g
Qualification tested to 0 to 2 kHz, 20 g sine
Limited only to those media which will not attack
polysulfone, silicon, fluorosilicone, silicone,
EPDM, and neoprene seals
24PC SERIES FLOW THROUGH ORDER GUIDE
Catalog
Listing
24PCE Type
24PCA Type
24PCB Type
24PCC Type
24PCD Type
24PCF Type
24PCG Type
Pressure
Range
psi
0.5
1.0
5.0
15
30
100
250
Span, mV
Min.
25
30
85
165
240
156
145
Typ.
35
45
115
225
330
225
212
Max.
45
60
145
285
420
294
280
Sensitivity
mV/psi
Typ.
70
45
23
15
11
2.25
0.85
Overpressure
psi Max.
20
20
20
45
60
200
500
Honeywell
Sensing and Control 1-800-537-6945 USA
1-815-235-6847 International 1-800-737-3360 Canada
21