Pre-amplified/5V Excitation/Gauge
Connection diagram
XFPM(NIBP) Data sheet
XFPM
Vcc
Vcc(DC5V)
Vout
Vout
GND
C2
C1
Remarks :
1)C1 : Connect 680pF capacitor within 2cm distance from leads.
2) C2 : Connect 0.01 F or more capacitor within 2cm distance
from leads.
GND
3) 4 and 5 pins must be No Connection. Open them as floating
ones completely, and do NOT connect to other line or each
other.
Transfer Function
Vout=Vs (P
+ ) (Pressure Error Temperature Error Multiplier
Vs)
Vs=5.0V
Notes ; Output, Vout, is not perfectly ratiometric power supply voltage Vs.
Model
XFPM-050KPGR-P1
pressure range
0 +300mmHg
0.00233
0.1
Pressure Error 1)(mmHg)
8.57
Temperature Error Multiplier=1
3
Temperature Error Multiplier
2
1
0
0
5
10
15
20
25
30
35
40
45
50
Temperature
Note ; Please read instruction “Notes” before using the sensor.
Fujikura reserves the right to change specifications without notice.
If you have any questions regarding technical issues or specifications, please contact us.
Sensor Engineering Department 5-1 Kiba 1-chome, Koto-ku, Tokyo 135-8512, Japan
Phone +81-(0)3-5606-1072 Fax. +81-(0)3-5606-1538
E-mail : sensor@fujikura.co.jp