90PF Slot Card
High Pass
Filter Options
90PF - H8E
8-Pole, 6-Zero
elliptic – 1.77
0.1 Hz to 300 kHz
±0.035
dB
0
±
0.2dB to 100kHz
0
±
0.5dB to 400kHz
500 kHz
(-6 dB) 1 MHz
Attenuation
80 dB
f
r
±
2% max.
-0.035 dB
-323.5°
±4º
max.
±2º
typ.
±2º
max.
±1º
typ.
0.56 f
r
0.60 f
r
1.00 f
r
2.00 f
r
Specifications
Transfer Function
90PF - H8B
8-Pole
Butterworth
0.1 Hz to 300 kHz
0.0 dB
0
±
0.2dB to 100kHz
0
±
0.5dB to 400kHz
500 kHz
(-6 dB) 1 MHz
Attenuation Rate
48 dB/octave
f
c
±
2% max.
-3 dB
-360°
±3º
max.
±1.5º
typ.
Tuning Range
Theoretical Transfer
Characteristics
Passband Ripple
(theoretical)
Pass Band Gain
(non-inverting)
Power Bandwidth
Small Signal
Bandwidth
Stop Band
Corner Frequency
Accuracy
Amplitude
Phase
Phase Match
1
f
c
to 100 kHz
f
r
– 1.25 f
r
1.25f
r
to 100 kHz
Filter Attenuation
(theoretical)
-83 dB
-48 dB
-3.01 dB
-0.01 dB
0.30 f
c
0.50 f
c
1.00 f
c
1.50 f
c
-80.0 dB
-60.0 dB
-0.03 dB
0.00 dB
Amplitude Accuracy
1
0.1 Hz to 100 kHz
(theoretical)
f
c
– 1.25 f
c
±0.3
dB max.
±0.15
dB typ.
1.25 f
c
– 400 kHz
±0.5
dB max.
f
r
– 1.25 f
r
±0.4
dB max.
±0.2
dB typ.
1.25 f
r
– 400 kHz
±0.5
dB max.
Total Harmonic
Distortion @ 1 kHz
Broad Band Noise
(5 Hz – 2 MHz)
<-100dB typ. @ 3V
rms
400µV
rms
typ.
<100nV/
Hz
typ.
<-100dB typ. @ 3V
rms
400µV
rms
typ.
<100nV/
Hz
typ
Narrow Band Noise
Note 1: Channel to channel match for the same transfer function set to the same frequency and operating configuration.
5
1784 Chessie Lane, Ottawa, IL 61350 • Tel: 800/252-7074, 815/434-7800 • FAX: 815/434-8176
e-mail: sales@freqdev.com
•
Web Address: http://www.freqdev.com