内容描述: 水位监测 [Water Level Monitoring]
分类和应用: 监控
文件页数/大小: 24 页 / 298 K
品牌: FREESCALE [ Freescale ]
 浏览型号AN1950的Datasheet PDF文件第1页浏览型号AN1950的Datasheet PDF文件第3页浏览型号AN1950的Datasheet PDF文件第4页浏览型号AN1950的Datasheet PDF文件第5页浏览型号AN1950的Datasheet PDF文件第6页浏览型号AN1950的Datasheet PDF文件第7页浏览型号AN1950的Datasheet PDF文件第8页浏览型号AN1950的Datasheet PDF文件第9页  
Table 1. MPXM2010D OPERATING CHARACTERISTICS (VS = 10 VDC, TA = 25°C unless otherwise noted, P1 > P2)  
Characteristic  
Symbol  
POP  
VS  
Min  
0
Typ  
Max  
10  
16  
Unit  
kPa  
Pressure Range  
Supply Voltage  
Supply Current  
Full Scale Span  
Offset  
10  
6.0  
25  
Vdc  
IO  
mAdc  
mV  
VFSS  
Voff  
24  
26  
1.0  
-1.0  
mV  
Sensitivity  
DV/DP  
2.5  
mV/kPa  
%VFSS  
Linearity  
-1.0  
1.0  
The amplified voltage signal from the positive sensor lead  
is VB. This amplification adds a small gain to ensure that the  
positive lead, V2, is always greater than the voltage output  
from the negative sensor lead, V4. This ensures the linearity  
of the differential voltage signal.  
Amplifier Induced Errors  
The sensor output needs to be amplified before being  
inputted directly to the microcontroller through an eight-bit A/D  
input pin. To determine the amplification requirements, the  
pressure sensor output characteristics and the 0-5 V input  
range for the A/D converter had to be considered.  
The amplification circuit uses three op-amps to add an  
offset and convert the differential output of the MPXM2010GS  
sensor to a ground-referenced, single-ended voltage in the  
range of 0–5.0 V.  
The pressure sensor has a possible offset of ±1 mV at the  
minimum rated pressure. To avoid a nonlinear response when  
a pressure sensor chosen for the system has a negative offset  
(VOFF), we added a 5.0 mV offset to the positive sensor output  
signal. This offset will remain the same regardless of the  
sensor output. Any additional offset the sensor or op-amp  
introduces is compensated for by software routines invoked  
when the initial system calibration is done.  
VB = (1+R7/R5) * V2 – (R7/R5) * VCC  
= (1+10/1000) * V2 + (10/1000)*(5.0 V)  
= (1.001) * V2 + 0.005 V  
The difference between the positive sensor voltage, VB,  
and the negative sensor voltage, VA is calculated and  
amplified with a resulting gain of 500.  
VC = (R12/R11) * (VB – VA)  
= (500 K/1K) * (VB – VA)  
= 500 * (VB – VA)  
The output voltage, VC, is connected to a voltage follower.  
Therefore, the resulting voltage, VC, is passed to an A/D pin of  
the microcontroller.  
To determine the gain required for the system, the  
maximum output voltage from the sensor for this application  
had to be determined. The maximum output voltage from the  
sensor is approximately 12.5 mV with a 5.0 V supply since the  
full-scale output of the sensor changes linearly with supply  
voltage. This system will have a maximum pressure of 4 kPa  
at 40 cm of water. At a 5.0 V supply, we will have a maximum  
sensor output of 5 mV at 4 kPa of pressure. To amplify the  
maximum sensor output to 5.0 V, the following gain is needed:  
The range of the A/D converter is 0 to 255 counts. However,  
the A/D Values that the system can achieve are dependent on  
the maximum and minimum system output values:  
Count = (VOUT – VRL) / ( VRH – VRL) x 255  
where VXdcr = Transducer Output Voltage  
VRH = Maximum A/D voltage  
VLH = Minimum A/D voltage  
Gain = (Max Output needed) / (Max Sensor Output  
and Initial Offset) = 5.0 V / (0.005 V + 0.005) = 500  
Count (0 mm H20) = (2.5 – 0) / (5.0 – 0) * 255 = 127  
Count (40 mm H20) = (5.0 – 0) / (5.0 – 0) * 255 = 255  
Total # counts = 255 – 127 = 127 counts.  
The gain for the system was set for 500 to avoid railing from  
possible offsets from the pressure sensor or the op-amp.  
The Voltage Outputs from the sensor are each connected  
to a non-inverting input of an op-amp. Each op-amp circuit has  
the same resistor ratio. The amplified voltage signal from the  
negative sensor lead is VA. The resulting voltage is calculated  
as follows:  
The resolution of the system is determined by the mm of  
water represented by each A/D count. As calculated above,  
the system has a span of 226 counts to represent water level  
up to and including 40 cm. Therefore, the resolution is:  
Resolution = mm of water / Total # counts  
VA = (1+R8/R6) * V4  
= (1+10/1000) * V4  
= (1.001) * V4  
= 400mm/127 counts = 3.1 mm per A/D count  