Relative mode  
Relative Zero allows the user to offset the meter by using the displayed  
value as the zero reference value. Practically all of the displayed  
readings can be set as the relative reference value including the  
MAX/MIN recording feature. Press the key momentarily to activate  
and to exit Relative Zero mode.  
Peak C capture mode  
Press PEAK key momentarily to activate the mode to capture voltage or  
current signal duration as short as 0.8ms. This mode is available in DC,  
AC, DC+AC modes of voltage and current functions. The LCD  
annunciators “C” & “MAX” turn on. The meter beeps when a new  
maximum or minimum reading is updated. Press the key momentarily to  
read throughout the Maximum (MAX), Minimum (MIN), and Maximum -  
Minimum readings. Press the key for 1 second or more to exit PEAK  
capture mode. Auto Power Off feature will be disabled automatically in  
this mode.  
Smart Power Auto Power Off  
Smart Power shuts the meter down automatically after approximately 17  
minutes of inactivity. The 17-minute time period is reset any time the  
rotary switch is moved or as long as a signal is being fed into the meter  
is greater than 10% of the range. This prevents auto powering off during  
long-term tests. To wake the meter, press the MAX/MIN key or turn the  
meter off and on again.  
Disabling Auto Power Off  
Press the RANGE key while turning the meter on to disable the Auto  
Power Off feature. Auto power off is also disabled in the MAX/MIN and  
Peak modes.  
Audible Beeper  
The audible beeper can be disabled by pressing the Hz key while  
turning the meter on.  
Backlighted display  
Press the SELECT key for 1 second or more to turn on or off the display  
backlight function. The backlight will automatically turn off after 30  
seconds.  
Hold H  
The hold function freezes the reading in the display. Press the HOLD H  
key momentarily to activate or to exit the hold function  