1.2 Resistance Measurement  
A re-configurable divider automatically provides a suitable full-scale range in resistance  
measurement mode. The following table summarizes the full-scale ranges and the  
reference resistors in each configuration. In 110.0MΩ, the maximum counting number  
is 1100 counts and the resolution is 0.1MΩ. So the full scale becomes 110.0MΩ.  
Configuration  
OR1  
Full Scale Range  
110.00Ω  
Divider Ratio  
Resister Connection  
R6  
R5  
100Ω  
1KΩ  
OR2  
1.1000KΩ  
OR3  
OR4  
11.000KΩ  
110.00KΩ  
R4  
R1R3  
10KΩ  
100KΩ  
OR5  
1.1000MΩ  
1MΩ  
R1R2  
R1  
OR6  
OR7  
11.000MΩ  
110.0MΩ  
10MΩ  
10MΩ  
R1  
1.3 Current measurement  
ES51916 has 2 automatic and 1 manual current measurement modes for multimeter and  
5 manual current measurement modes for clampmeter. The following table summarizes  
the full-scale range of each mode. When ES51916 operate in the current measurement  
modes for multimeter, It takes high input from pin IVSH or IVSL, low input from pin  
SGND and reference voltage from pin VR. When ES51916 operate in the current  
measurement modes for Clampmeter, It takes high input from pin ADP, low input from  
pin SGND and reference voltage from pins VA+ and VA-. The test circuits are shown in  
page 27-28.  
Mode  
Range Selection  
Full Scale  
110.00µA / 1100.0µA  
11.000mA / 110.00mA  
11.000A  
Designed for  
Multimeter  
Multimeter  
Multimeter  
Automatic Mode 1  
Automatic Mode 2  
Manual Mode 1  
µA  
mA  
A
Manual Mode 2  
Manual Mode 3  
Manual Mode 4  
Manual Mode 5  
Manual Mode 6  
A
A
A
A
A
1.1000A  
11.000A  
110.00A  
1100.0A  
11000A  
Clampmeter  
Clampmeter  
Clampmeter  
Clampmeter  
Clampmeter  
Note:  
1. In Manual DC current measurement modes for clampmeter, ES51916 provides Zero function for offset  
removing.  
2. In Automatic mode1 and Automatic mode2, the pin VBAR is used to control the display of “µ” and  
“m” sign on the LCD panel. If pin VBAR is floating or connected to DGND, the “µ” and “m” sign will  
be displayed. If pin VBAR is connected to V-, the “µ” and “m” sign will not be displayed.  
v1.2  
9
11/09/06  
 复制成功!