ES51927
6000 Counts Smart DMM
3.2 Range function
RANGE pin switches to and from automatic and manual mode, and while in
manual mode, changes the full-scale range. The following figure shows the state
transition.
1 push < 1 sec
Automatic
Mode
Manual
Mode
1 push > 1 sec
Range up
1 push < 1 sec
3.3 Hold and delay-hold function
The data hold mode makes the meter stop updating the LCD panel. This mode
can be nested in most of the special modes. In auto scanning mode, the HOLD
key is not available. Enabling hold function in automatic mode makes the meter
switch to manual mode. Hold function can be cancelled by changing the
measurement mode, pressing RANGE, or push HOLD again.
The delay-hold function is enabled when HOLD key is pressed larger than 2
seconds. When delay-hold function is entered successfully, the meter will stop to
update the LCD data after six seconds delayed. During the six seconds waiting,
the HOLD symbol on LCD panel will be blinking.
Push > 2s
Push < 2s
Push < 2s
Normal mode
(Auto/Manual)
HOLD mode
(Manual)
6sec past
Delay State
Push > 2s
Push < 2s
(Auto/Manual)
Ver 2.9
14
10/06/25