ES51982
3 1/2 CAP AUTO DMM
5.4 Operating Timing
ES51982 incorporates a dual slope ADC with four phases: ZI, AZ, INT and DINT.
The timing of each phase is listed below.
*Voltage /ohm/ADP measurement:
Phase
ZI
AZ
INT
DINT
Time
100ms
100ms
100ms
200ms
Ps. In the voltage and ohm measurement with auto mode, if the range is changed,
the internal clock rate will increase ten times and the new measurement cycle
becomes 1/10 times of the original cycle until the range is stable.
*Current mode
Phase
ZI
AZ
INT
DINT
Time
100ms
100ms
100ms
100ms
*Continuity/diode measurement
Phase
ZI
AZ
INT
DINT
Time
10ms
10ms
10ms
20ms
*Capacitance measurement:
Range
2.000nF
20.00nF
200.0nF
2.000uF
20.00uF
200.0uF
2.000mF
Total Measurement Time
0.5sec
0.5sec
0.5sec
0.5sec
0.5sec
1sec
2.5sec
*.Frequency measurement
one measurement cycle = 1.05 sec
ps. In the frequency measurement with auto mode, if the range is changed, the
internal clock rate will increase ten times and the new measurement cycle
becomes 1/10 times of the original cycle until the range is stable.
18
11/01/05