ES51969  
6,600 Counts Auto DMM  
1.8 Frequency counter  
The time base of the frequency counter is derived from an external crystal oscillator by  
4,000,000  
Tcounter  
=
Fosc  
where Fosc is the frequency of the crystal oscillator. Thus, the counter has a 1-second  
time base when a 4MHz oscillator is used. The frequency counter can select the proper  
range automatically or manually. Auto-range operation extends over six decades, from  
660.0Hz to 66.00MHz. The following table summarizes the full-scale range of the  
frequency counter.  
Range  
FR1*  
FR2  
FR3  
FR4  
Full Scale  
660.0Hz  
6.600KHz  
66.00KHz  
660.0KHz  
6.600MHz  
66.00MHz  
FR5  
FR6  
*If input frequency is less than 10.0Hz, ES51969 will show 00.0Hz.  
1.9 Duty cycle measurement  
In frequency mode, press the VAHZ bottom will force ES51969 enter duty cycle  
measurement mode. The duty cycle measurement mode takes high input from FREQ  
pin, and low input from SGND pin. The annunciator “%” on LCD panel indicates  
whether the duty cycle measurement mode is activated. When the duty cycle is less  
than 10.0%, the LCD panel shows UL and the UL bit of RS232 data becomes high.  
When the duty cycle is more than 94.9%, the LCD panel shows OL and the OL bit of  
RS232 data becomes high. If the frequency of input signal is less than 10.0Hz, the LCD  
panel also shows UL and the UL bit of RS232 data becomes high. ES51969 also  
provides a single duty cycle mode, which FC1~FC4 = {1,1,1,0}, as an independent  
rotary switch for the meters without a frequency mode configured.  
1.10 Temperature measurement mode  
Temperature measurement mode takes input signal from ADP pin and reference voltage  
from (VA+ - VA-). ES51969 has a built-in ºC-to-ºF scale translation circuit, and only  
needs an external ºC scale application circuit. The application circuit for ºF scale is not  
required. In temperature measurement mode, the default range is 6600 ºC or ºF. The  
VBAR pin is used to control the first decimal point (DP1) on the LCD panel. When  
VBAR is pulled to V-, DP1 will be turned on and the full scale becomes 660.0 ºC or ºF.  
v2.6  