3 3/4 MANU DMM
present LOGIC state
LOGIC high level setting
buzzer on level
ES51961
The default LOGIC mode panel
LOGIC probe input high should connect to PKIN pin, the LOGIC probe
input low should connect to V- pin (not VIN- pin), so the maximum voltage
difference between LOGIC high & low can be 6V.
Example:
The above panel means now is logic high (the input > 4.0V) and the
buzzer turn on.
The above panel means now is logic unknown (the input between
0.8V~2.4V) and once the LOGIC level change to high (>2.4V), the
buzzer will turn on.
9
2006/1/24