ES51961
3 3/4 MANU DMM
Decimal point select (Under voltage mode)
DP2 DP1 Decimal point
H
H
L
L
H
L
H
L
None
Point1
Point2
Point3
Auto-Power-Off function
Once power on the ES51961, the auto-power-off function will be enabled
and if the functions are not changed in 30 minutes, the auto-power-off condition
will be happened. When power-off happens, the final data is saved. If continue
to push anyone of the push function (except HOLD pin) then power on the
ES51961, the auto-power-off function will be disabled, and LCD segment
"APO" will be turn off.
Re-Power-On
Once auto-power-off happens, push anyone of the push function or change
the rotary mode can re-power-on the ES51961. When anyone of the push
function is enabled, the storage value will display with HOLD mold (except
logic state), if auto-power-off happened under MAX/MIN (or PEAK, REL)
mode, the PMAX , PMIN , or relative and reference value will be keeped. But if re-
power-on the ES51961 with rotary mode, the storage value will be clear.
SLEEP
When auto-power-off happens, this pin output will change from V+ to V-,
once re-power on happen, this pin output will change from V- to V+.
Buzzer
(1) The BUZZER turns on in the following condition :
•1 beep: Measurement function changed ,power and re-power on ,HOLD
MAX/MIN, REL, PEAK, BKLIT, SET, RANGE or ZOOM are
17
2006/1/24