ASM200
200W High Power Density Medical Grade Open Frame Power Supplies
BSM200 series
ENVIRONMENTAL SPECIFICATIONS
EMC CERTIFICATIONS
Conducted Emissions
Radiated Emissions
Oper. Temperature Range
Storage Temperature Range
Humidity
-20 to +40°C at Full Load
EN60601-1-2 Class B
EN55011 Class B
EN60601-1-2 Class B
EN55011 Class B
EN61000-4-2 Criteria A
Level 3
-40 to +85°C
0 to 95%, non-condensing
0 to 10,000 ft., 0 to 3048 m
30G pk. Half sine, 6 axis
2 G RMS, 5 Hz to 500 Hz
3 axis, 30 min
Altitude
ESD Susceptibility
Air Discharge
ESD Susceptibility
Contact Discharge
Radiated
Shock
Vibration
EN61000-4-2 Criteria A
Level 2
EN61000-4-3 Criteria A
Level 2
SAFETY CERTIFICATIONS
Susceptibility
EFT/Burst
EN61000-4-4 Criteria A
Level 3
UL/cUL
UL EU
ANSI/AAMI ES60601-1 3rd
Edition
Surge
EN61000-4-5 Criteria A
Level 2
IEC 60601-1 3rd Edition
CB Reporting Scheme
Conducted
EN61000-4-6 Criteria A
Level 2
Susceptibility
EMC Certification Notes:
All specifications are typical at nominal input, full load, and 25°C unless
specifiied otherwise
1) Emissions compliance at both 115V and 230V
2) For full EMC compliance, 4 mounting holes should be attached to an
electrically conductive surface.
OUTPUT POWER DERATING vs. AMBIENT TEMPERATURE
Output Power vs. Ambient
Temperature, 90-264 VAC,
Natural Convection
Derate 4.28W/°C
from 40°C to 75°C
©2015 AstrodyneTDI Corporation
ASM/BSM200S rev. B 07-1-15