50W 3rd Edition Medical Grade Desktop Switching Power Adapters
AMP510 series
INPUT SPECIFICATIONS
GENERAL SPECIFICATIONS
Input Voltage Range
Frequency Range
90-264 VAC
47-63 Hz
1.35A
Isolation
I/P-O/P: 4000VAC
I/P-Ground: 1500VAC
Input Current Lowline
Input Current Highline
Inrush Current, typ (Cold Start)
115VAC Input
Isolation Resistance
Efficiency
50MΩ
0.8A
See Selection Charts
Safety
ANSI/AAMI ES 60601-1: 2005 (UL/cUL 3rd Edition)
EN 60601-1:2006 (TUV/T-mark 3rd Edition)
12 Amps (15Amps Max.) *
26 Amps (30 Amps Max.) *
0.1mA typ. @ 240VAC
See Selection Charts
230VAC Input
Leakage Current
PHYSICAL SPECIFICATIONS
No-Load Power Consumtion
Size
5.75” x 2.99” x 1.73”
Desktop Style
Construction
Weight
OUTPUT SPECIFICATIONS
19.75 oz (560g)
Voltage and Current
See Selection Charts
0-FL Single Outputs
10% Multi-Outputs
Minimum Load Required
ENVIRONMENTAL SPECIFICATIONS
Load Regulation
See Selection Charts
+/- 0.5%, typ (1% Max.)
+/-0.04%/°C, typ
Oper. Temperature
0 to +70°C
Line Regulation
See Derate Curve
-40 to +85°C *
0 to + 95%*
Vin = 220VAC
Vin = 110VAC
Temperature Coefficient
Ripple/Noise
Storage Temperature
Relative Humidity
EMC
0.5% Pk-Pk, typ (1% Max.), 2% 3.3V
112-132%*
110-150% of Rated Current
Over Voltage Protection
Over Current Protection
Transient Response
Start Up Time (min./typ./max.)
Hold Up Time
CISPR-11 B
FCC Part 18 B
50% to 100% Load
2 Seconds max.
4mS
MTBF (MIL-HDBK-217F, 25°C) 100,000 Hours
16mS min. at 110VAC Input
OUTPUT DERATING CURVE
All specifications are typical at nominal input, full load, and 25°C unless other-
wise noted
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
WWW.ASTRODYNETDI.COM