15W Medical Grade Desktop Switching Power Adapters  
AMP1501 series  
INPUT SPECIFICATIONS  
GENERAL SPECIFICATIONS  
Input Voltage Range  
Frequency Range  
Input Current Max.  
90-264 VAC (Universal)  
47-63 Hz  
Input-Out Isolation  
Isolation Resistance  
Efficiency (FL, 230Vin)  
Safety  
5656VDC (4000VAC)  
50 MΩ min.  
Low Line (115VAC)  
High Line (230VAC)  
0.3W @ 230VAC  
0.33A  
75%, typ  
0.18A  
(c)UL60601-1, EN60601-1  
No Load Power Consumption  
Inrush Current, typ (Cold Start) Low Line (115VAC)  
High Line (230VAC)  
16A max.  
30A max.  
PHYSICAL SPECIFICATIONS  
Leakage Current  
0.3mA MAX.  
Size  
4.11” x 1.65” x 1.22”  
Desktop Style  
@ 240 VAC, FL  
Construction  
Weight  
6.0oz (170g)  
OUTPUT SPECIFICATIONS  
ENVIRONMENTAL SPECIFICATIONS  
Voltage and Current  
Output Power Range  
Load Regulation (230Vin)  
Line Regulation  
See Selection Charts  
0 to FL Rating  
± 5%, max.  
Oper. Temperature  
0 to +50°C (FL) *  
50% Load @ 70°C  
(See Derate Curve)  
-40 to +85°C *  
5% to 95%, non-cond *  
FCC Part 15 Class B  
EN55022B  
± 1%, max.  
Transient Response  
Temperature Coefficient  
Ripple/Noise  
100Vin, FL to 1/2L  
±0.04%/°C, typ  
FL, 90Vin  
4mS  
Storage Temperature  
Relative Humidity  
1% Pk-Pk  
EMC  
110Vin:  
Over Voltage Protection  
Over Current Protection  
Short Circuit Protection  
Thermal Shutdown  
112-132%  
110-150%  
MTBF  
100,000 Hrs  
Continuous, self-recovering *  
-20 to 125°C  
Efficiency Level  
V
Start Up / Hold Up Time  
100Vin, FL  
1S max. / 10 mS min.  
DERATE CURVE  
100  
80  
60  
40  
20  
0
10  
20  
30  
40  
50  
60  
70  
All specifications are typical at nominal input, full load, and 25°C unless other-  
wise noted  
* These are stress ratings. Exposure of the devices to any of these conditions  
may adversely affect long term reliability. Proper operation under conditions  
other than the standard operating conditions is neither warranteed nor implied.  
ASTRODYNE USA: 1-800-823-8082  
ASTRODYNE PACIFIC: 886-2-26983458  
WWW.ASTRODYNETDI.COM