12W Medical Grade Wallmount Switching Power Adapters  
AMP1201 series  
PHYSICAL SPECIFICATIONS  
INPUT SPECIFICATIONS  
Size  
2.56” x 2.17” x 1.40”  
(65x55x35.6mm)  
Wallmount Style  
5.64oz (160g)  
Input Voltage Range  
Frequency Range  
Input Current  
80-275 VAC (100-240V Nominal)  
47-63 Hz  
Construction  
Weight  
Low Line (100VAC)  
0.29A max.  
High Line (240VAC)  
0.16A max.  
50A max.  
Inrush Current, typ (Cold Start) Low Line (100VAC)  
High Line (230VAC)  
100A max.  
ENVIRONMENTAL SPECIFICATIONS  
Leakage Current  
<100 uA  
Oper. Temperature  
-10 to +50°C (FL) *  
Protection Class  
Double Insulated, Class II  
50% Load @ 70°C  
No Load Power Consumption  
0.3W  
(See Derate Curve Next Page)  
Free Air Convection  
Cooling  
OUTPUT SPECIFICATIONS  
Flammability Rating  
Storage Temperature  
Operating/Storage Humidity  
EMC  
UL94V-1  
-40 to +85°C (10-95% RH) *  
0% to 95%, non-cond *  
Compliance to EN55011 (CISPR11),  
EN61000-3-2,-3 Class B  
Voltage and Current  
See Selection Chart  
0 to FL Rating  
± 1%, max.  
Output Power Range  
Line Regulation (FL) (Note 1)  
Over Voltage Protection  
Over Load Protection  
Transient Response  
112-132%  
MTBF (MIL-HDBK-217F, 25 °C) 100,000 Hrs min.  
110-150% Auto-Recovery  
110Vin, FL  
Efficiency Level  
V
4mS  
Surge Voltage  
2kV min.  
Temperature Coefficient  
Ripple/Noise (Note 3)  
Short Circuit Protection  
Start Up / Hold Up Time  
±0.04%/°C, typ  
ESD  
Air:  
IEC61000-4-2, 8kV min.  
IEC61000-4-2, 6kV min.  
3000M max.  
FL, 90Vin  
1% Pk-Pk max.  
Contact:  
Continuous, self-recovering *  
100Vin, FL  
Operating Altitude  
2S max. / 16mS min.  
Vibration  
10 ~ 500Hz, 10min./1cycle, 60min.  
each along X, Y, Z axes, 5G min  
GENERAL SPECIFICATIONS  
NOTES  
Input-Out Isolation  
Efficiency(FL, Nom. Vin)  
Safety  
4000 VAC  
See Selection Chart  
1. Line Regulation defined as changing the input voltage ±10% from  
nominal line at rated load.  
ANSI/AAMI ES 60601-1:2005  
(UL/cUL/EN/IEC 3rd Edition)  
2. Total Regulation is defined by as changing the load ±40% from 60%  
rated load.  
3. Ripple & Noise is measured by using a 20MHz bandwidth limited  
oscilloscope and terminating the output with a 0.47uF capacitor at  
rated load and nominal line.  
All specifications are typical at nominal input, full load, and 25°C unless other-  
wise noted  
* These are stress ratings. Exposure of the devices to any of these conditions  
may adversely affect long term reliability. Proper operation under conditions  
other than the standard operating conditions is neither warranteed nor implied.  