63W Medical Power Adapters for Health Care Applications
AHP6301 series
INPUT SPECIFICATIONS
PHYSICAL SPECIFICATIONS
Input Voltage Range
Frequency Range
Input Current
80-275 VAC (100-240 Nominal)
47-63 Hz
Size
5.17” x 2.18” x 1.46”
(131 x 55.4 x 37mm)
Desktop Style
Low Line (100VAC)
High Line (240VAC)
1.62A max.
Construction
Weight
0.72A max.
37A max.
74A max.
13.4oz (380g)
Inrush Current, typ (Cold Start) Low Line (100VAC)
High Line (240VAC)
Flammability Rating
UL94V-1
Leakage Current (240 Vin)
0.1mA
ENVIRONMENTAL SPECIFICATIONS
No Load Power Consumption
0.5 Watts
Oper. Temperature
-10 to +40°C (FL) *
50% Load @ 70°C
OUTPUT SPECIFICATIONS
(See Derate Curve Last Page)
Free Air Convection
Voltage and Current
Output Power Range
Load Regulation
See Selection Charts
Cooling
0 to FL Rating
Storage Temperature
Operating Humidity
EMC
-40 to +85°C, 10~95% RH *
0% to 95%, non-cond *
Compliance to EN55011 (CISPR11),
EN61000-3-2, -3 Class B
100,000 Hrs
±5%, max. 60%L, changing ±10%
±1%, max. changing line ±10%
110-150%, Auto-Recovery
112-132%
Line Regulation (FL)
Over Load Protection*
Over Voltage Protection*
Short Circuit Protection
Transient Response
Temperature Coefficient
Ripple/Noise (Note 1)
Hold Up Time (Note 2)
Start Up Time
MTBF (MIL-HDBK-217F)
Surge Voltage
Auto Recovery
2kV
110Vin, FL
4mS
ESD - Air
IEC61000-4-2, 8kV
±0.04%/°C, typ
100mV Pk-Pk max.
100~240Vin, FL
ESD - Contact
IEC61000-4-2, 6kV
Operating Altitude (Elevation)
Vibration
3000M
12mS min.
5G, 10 ~ 500Hz, 10min./1cycle,
60min. each along X, Y, Z axes
120cm, 1 ea. on six sides
100~240Vin, FL 0.3S typ., 2S max.
Drop Test
GENERAL SPECIFICATIONS
NOTES
Input-Out Isolation
Input-GND
4000 VAC, Limit Current <10mA
2800 VDC
1. Ripple & noise is measured by using 20MHz bandwidth limited
oscilloscope and terminated each output with a 0.47uF capacitor at
rated load and nominal line.
Isolation Resistance
Efficiency
50MΩ
(FL, 230Vin) See Selection Chart
ANSI/AAMI ES 60601-1:2005
(UL/cUL 3rd Edition),
EN 60601-1:2006
Safety
2. Hold up time is measured from the end of the last charging pulse to
the time which the main output drops down to low limit of main output at
rated load and nominal line.
(TUV/T-mark 3rd Edition)
AHP6301A, 6301C: Class I
AHP6301B: Double Insulated
Class II
Protection Class
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
All specifications are typical at nominal input, full load, and 25°C unless other-
wise noted
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458
WWW.ASTRODYNETDI.COM