30W Medical Grade Switching Power Supplies
AHP3201 series
INPUT SPECIFICATIONS
PHYSICAL SPECIFICATIONS
Input Voltage Range
Frequency Range
Input Current
80-275 VAC (100-240 Nominal)
47-63 Hz
Size
4.21” x 1.85” x 1.17”
(107 x 47 x 29.6mm)
Desktop Style
Low Line (100VAC)
High Line (240VAC)
0.6A max.
Construction
Weight
0.4A max.
28A max.
67A max.
9.87 oz (280g)
UL94V-1
Inrush Current, typ (Cold Start) Low Line (100VAC)
High Line (240VAC)
Flammability Rating
No Load Power Consumption
0.3 Watts
ENVIRONMENTAL SPECIFICATIONS
Oper. Temperature
-10 to +40°C (FL) *
50% Load @ 70°C
OUTPUT SPECIFICATIONS
Voltage and Current
Output Power Range
Total Regulation
See Selection Chart
0 to FL Rating
(See Derate Curve Last Page)
Free Air Convection
Cooling
See Selection Chart
±1%, max.
Storage Temperature
Operating Humidity
EMC
-40 to +85°C, 10~95% RH *
0% to 95%, non-cond *
Compliance to EN55011 (CISPR11),
EN61000-3-2, -3 Class B
100,000 Hrs
Line Regulation (FL) (Note 3)
Over Load Protection*
Over Voltage Protection*
Short Circuit Protection
Transient Response
Temperature Coefficient
Ripple/Noise (Note 1)
Hold Up Time (Note 2)
Start Up Time
110-150%, Auto-Recovery
112-132%
Auto Recovery
110Vin, FL
MTBF (MIL-HDBK-217F)
Surge Voltage
4mS
2kV
±0.04%/°C, typ
See Selection Chart
100~240Vin, FL
ESD - Air
IEC61000-4-2, 8kV
ESD - Contact
IEC61000-4-2, 6kV
12mS min.
Operating Altitude (Elevation)
Vibration
3000M
100~240Vin, FL 0.3S typ., 2S max.
5G, 10 ~ 500Hz, 10min./1cycle,
60min. each along X, Y, Z axes
GENERAL SPECIFICATIONS
NOTES
Input-Out Isolation
Input-GND
4000 VAC, Limit Current <10mA
2800 VDC
1. Ripple & noise is measured by using 20MHz bandwidth limited
oscilloscope and terminated each output with a 0.47uF capacitor at
rated load and nominal line.
Isolation Resistance
Efficiency
50MΩ
(FL, 230Vin) See Selection Chart
ANSI/AAMI ES 60601-1:2005
(UL/cUL 3rd Edition),
EN 60601-1:2006
Safety
2. Hold up time is measured from the end of the last charging pulse to
the time which the main output drops down to low limit of main output at
rated load and nominal line.
3. Full Load, Vin=100~120VAC or 200~240VAC
(TUV/T-mark 3rd Edition)
Class II
Class
* These are stress ratings. Exposure of the devices to any of these conditions
may adversely affect long term reliability. Proper operation under conditions
other than the standard operating conditions is neither warranteed nor implied.
All specifications are typical at nominal input, full load, and 25°C unless other-
wise noted
ASTRODYNE USA: 1-800-823-8082
ASTRODYNE PACIFIC: 886-2-26983458