欢迎访问ic37.com |
会员登录 免费注册
发布采购

LPS44-M 参数 Datasheet PDF下载

LPS44-M图片预览
型号: LPS44-M
PDF下载: 下载PDF文件 查看货源
内容描述: 40瓦医疗 [40 Watts Medical]
分类和应用: 医疗
文件页数/大小: 3 页 / 83 K
品牌: ASTEC [ Astec America, Inc ]
 浏览型号LPS44-M的Datasheet PDF文件第2页浏览型号LPS44-M的Datasheet PDF文件第3页  
Embedded Power for  
Business-Critical Continuity  
Rev. 04.14.06  
LPS40-M Series  
1 of 3  
LPS40-M Series  
40 Watts  
RoHS  
Medical  
Total Power:  
40-55 Watts  
Input Voltage: 85-264 VAC  
120-300 VDC  
# of Outputs:  
Triple  
Electrical Specifications  
Input  
Special Features  
• Medical safety approvals  
• Universal input  
• 3" x 5" footprint  
• Less than 1U high  
Input range  
Frequency  
Inrush current  
Input current  
Efficiency  
85-264 VAC; 120-300 VDC  
47-440 Hz  
<18A peak @ 115VAC; <36A peak @ 230VAC, cold start @ 25°C  
1A max. (RMS) @ 115VAC  
70% typical at full load  
• Remote sense on main output  
• Low output ripple  
EMI filter  
FCC Class B conducted; CISPR 22 Class A conducted;  
EN55022 Class A conducted; VDE 0878 PT3 Class A conducted  
<75μA @ 50/60 Hz, 264 VAC input  
• Adjustable 5V output  
• Overvoltage protection  
• Overload protection  
• 110 KHz switching frequency  
• LPX40 enclosure kit available  
Safety ground  
leakage current  
Output  
Maximum power  
Adjustment range  
Hold-up time  
40W for convection; 55W with 30CFM forced air  
-5, +10% minimum  
20ms @ 40W load, 115VAC nominal line  
Overload protection  
Short circuit protection on all outputs  
Case overload protected @ 110-145% above peak rating  
Overvoltage protection 5V output; 5.7 to 6.7VDC. Other outputs 10% to 25%  
above nominal output  
Remote sense  
Compensates for 0.5 V lead drop minimum.  
Will operate without remote sense connected.  
Reverse connection protected.  
Safety  
VDE  
UL  
0750/EN60601-1 (IEC601)  
UL2601 E182560  
CSA 22.2 No.601.1  
LR109492  
CSA  
CE Mark (LVD)