10  
Main unit specifications  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,  
approx. 300 g (10.6 oz) Accessories: None  
REC & MEM functions (version 2.00 or later. Real-time save function: version 2.30 or later)  
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up  
VOLTAGE/TEMP UNIT 8937  
Measurement functions  
time and zero-adjust; accuracy guaranteed for 1 year)  
20ms to 1hour/DIV; 16 settings, 1 DIV = 100 samples, time axis  
compression 1/2 to 1/500, 8 settings  
Note: Sampling period 1/100 of time axis range at memory recorder function  
Time axis  
(REC)  
Number of channels: 2, for voltage measurement/temperature measurement with thermocouple  
Voltage input: metallic BNC connector (input impedance 1MΩ, input capacitance 50pF)  
thermocouple input: terminal connector (input impedance min. 5.1MΩ), Max. rated  
,
100μs to 5 minutes/DIV; 20 settings, 1 DIV = 100 samples, time axis  
zoom ×2 to ×10; 3 settings, compression 1/2 to 1/10,000, 12 settings  
Note: Sampling period 1/100 of time axis range (min. 1μs)  
Input connectors  
Time axis  
(MEM)  
voltage to earth: 30Vrms or 60V DC (with input isolated from the unit, the maximum voltage  
that can be applied between input channel and chassis and between input channels without damage)  
500μV to 2V/DIV, 12 ranges, full scale: 20DIV, low-pass filter: 5Hz/  
500Hz/ 5kHz/ 100kHz, Measurement resolution: 1/80 of measurement  
range (using 12-bit A/D conversion; installed in the 8841/8842)  
REC: Settable in 1-division steps, 25 to 1,000 DIV (5,000 DIV at  
expanded memory), or continuous  
MEM: Settable in 1-division steps, 25 to 2,000 DIV (10,000 DIV at  
Voltage measurement  
range  
Recording length  
expanded memory)  
10°C to 100°C/DIV, 4 ranges, full scale: 20DIV, low-pass filter:  
5Hz/500Hz, Measurement resolution:1/80 of measurement range  
(using 12-bit A/D conversion; installed in the 8841/8842)  
Temperature  
measurement range  
REC: timer trigger, or OFF  
Trigger source  
MEM: CH1 to CH16 (analog), logic A to D, or external trigger  
Time axis: 50ms/DIV to 5 minutes/DIV (less than 8 channels)  
100ms/DIV to 5 minutes/DIV (more than 9 channels) (1 DIV=100  
samples, sampling period 1/100 of time axis range)  
,
K: -200 to 1350°C, E: -200 to 800°C, J: -200 to 1100°C, T: -200 to 400°C,  
N: -200 to 1300°C, R: 0 to 1700°C, S: 0 to 1700°C, B: 300 to 1800°C,  
Reference junction compensation: internal/ external (switchable)  
Thermocouple  
range  
Real-time save  
functions  
Note: Used with the PC card,  
Save destination: PC card, or MO disk in the optional MO UNIT 9607  
Trigger: trigger settings are not applicable to the MEM waveform,  
and the start of MEM waveform recording coincides with the start  
of the REC waveform.  
Voltage input: 1MS/s, Temperature measurement: 4kS/s  
(simultaneous sampling in 2 channels)  
Highest sampling rate  
Voltage input: DC amplitude ±0.4% of full scale, zero position ±0.15% of full  
scale, Temperature measurement (K, E, J, T, N): ±0.1% of full scale ±1°C,  
±0.1% of full scale ±2°C (-200 to 0°C), (R, S): ±0.1% of full scale ±3°C, (B):  
±0.1% of full scale ±4°C (400 to 1800°C) , Reference junction compensation  
accuracy: ±0.1% of full scale ±1.5 °C (internal reference junction compensation)  
firmware version 2.60 or later  
Time axis at REC: 200ms/DIV to 1 hour/DIV (less than 8 channels)  
,
Accuracy  
500ms/DIV to 1 hour/DIV (more than 9 channels) (1 DIV=100 samples,  
sampling period 1/100 of time axis range)  
Only REC waveform is output when printer output is started,  
reprinting of stored REC waveform data (last 1,000 DIV; or 5,000  
DIV at expanded memory), Additional recording function (recording is  
resumed without overwriting previous data), variable display  
Frequency  
characteristics  
Voltage input: DC to 400 kHz +1/3dB  
Temperature measurement: DC to 1kHz +1/3dB  
Other functions  
DC, GND, AC  
Input coupling  
30Vrms or 60V DC (the maximum voltage that can be applied across input pins without damage)  
Max. allowable input  
FFT functions (version 2.00 or later.)  
Storage waveform, Linear spectrum, RMS spectrum, Power  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,  
approx. 250 g (8.8 oz) Accessories: Conversion cable × 2  
spectrum, Cross-power spectrum, Auto-correlation function,  
Histogram, Transfer function, Cross-correlation function, Unit-  
impulse response, Coherence function, Octave analysis  
Analysis mode  
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 1 hour of warm-up  
STRAIN UNIT 8939  
time and auto-balance; accuracy guaranteed for 1 year)  
1 or 2 selected channels out of all analog channels  
Analysis channels  
Frequency range  
Measurement  
functions  
Number of channels: 2, for distortion measurement (electronic  
auto-balancing, balance adjustment range within ±10000με)  
133mHz to 400kHz, External, (resolution 1/400, 1/800, 1/2000, 1/4000)  
Via conversion cable, TAJIMI PRC03-12A10-7M10.5, Max. rated voltage to  
earth: 30Vrms or 60V DC (with input isolated from the unit, the maximum voltage that can  
be applied between input channel and chassis and between input channels without damage)  
Number of  
sampling points  
1000, 2000, 5000, 10000 points  
Input connectors  
Rectangular, Hanning, Exponential  
Windows  
Strain gauge converter, bridge impedance: 120Ω to 1kΩ, bridge  
voltage 2 ±0.05V  
Suitable  
transducer  
Time axis / frequency axis simple averaging, exponential averaging,  
peak hold  
Averaging  
function  
20με to 1000με/DIV, 6 ranges, full scale: 20DIV, low-pass filter:  
10Hz/30Hz/300Hz/3kHz  
Measurement  
range  
PC Software Specifications  
1/80 of measurement range (using 12-bit A/D conversion; installed in 8841/8842)  
Measurement resolution  
Highest sampling rate  
Wave Viewer (Wv) Software (Application disk CD-R, bundled accessory)  
1MS/s (simultaneous sampling in 2 channels)  
• Simple display of waveform file  
• Text conversion: convert binary data file to text format, with  
selectable space or tab separators in addition to CSV, and speci-  
Accuracy After auto-balancing DC amplitude: ±(0.5% of full scale +2με), zero position: ±0.5% of full scale  
DC to 20 kHz +1/3dB  
Frequency characteristics  
Max. allowable input  
fiable section, thinning available  
• Display format settings: scroll functions, enlarge/reduce dis-  
Functions  
10V DC + AC peak (the maximum voltage that can be applied across input pins without damage)  
play, display channel settings  
• Others: voltage value trace function, jump to cursor/trigger  
position function  
CONVERSION CABLE 9318 (to connect 9270 to 9272, 9277 to 9279 and 8940)  
CONVERSION CABLE 9319 (to connect 3273, 3273-50 and 8940)  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,  
approx. 300 g (10.6 oz) Accessories: None  
Windows 95/98/Me, Windows NT 4.0 (SP3 or later), 2000, XP  
Compatible PC operating systems  
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up  
F/V UNIT 8940  
time and zero-adjust; accuracy guaranteed for 1 year)  
Number of channels: 2, for voltage input based frequency measurement, integration,  
pulse duty ratio, current (with optional clamp-on sensor), and voltage measurement  
Input unit specifications (sold separately)  
Measurement  
functions  
Metallic BNC connector (input impedance 1MΩ, input capacitance 60pF), sensor  
connector (dedicated connector for clamp-on sensor via conversion cable, common  
ground with recorder), Max. rated voltage to earth: 30Vrms or 60V DC (with  
input isolated from the unit, the maximum voltage that can be applied between input  
Dimensions and mass: approx. 170 (6.69in) W × 20 (0.79in) H × 148.5 (5.85in) D mm,  
approx. 290 g (10.2 oz) Accessories: None  
Input connectors  
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up  
time and zero-adjust; accuracy guaranteed for 1 year)  
ANALOG UNIT 8936  
channel and chassis and between input channels without damage)  
Compatible current sensors 9270, 9271, 9272, 9277, 9278, 9279, 3273, 3273-50  
Number of channels: 2, for voltage measurement  
Measurement functions  
Frequency: DC to 100kHz, with 0.05Hz to 5kHz/DIV, 11 ranges, 5  
(r/min) to 500 (r/min)/DIV, 5ranges, P50Hz (40 to 60Hz), P60Hz (50 to  
70Hz) * Power line frequency measurement requires the DIFFERENTIAL PROBE 9322 or PT  
Isolated BNC connector (input impedance 1MΩ, input capacitance 30pF), Max. rated  
voltage to earth: 370V AC, DC (with input isolated from the unit, the maximum voltage that  
can be applied between input channel and chassis and between input channels without damage)  
Input connectors  
9303, Accuracy: ±0.2% of full scale (except 10kHz/DIV range), ±0.7% of  
full scale (10kHz/DIV range), ±0.032Hz (P50Hz, P60Hz range)  
5mV to 20V/DIV, 12 ranges, full scale: 20DIV, AC voltage for  
possible measurement/display using the memory function: 280V  
rms, low-pass filter: 5Hz/500Hz/5kHz/100kHz  
Measurement  
range  
Measurement  
Integration: DC to 90kHz, with 5counts to 500kcounts/DIV, 11 ranges  
range  
Pulse duty ratio: 10Hz to 100kHz, with 100% of full scale, 1 range,  
Accuracy: ±1% of full scale (10Hz to 10kHz)  
Threshold: 10 to +10V (settable in 0.2V steps)  
Full scale: 10DIV, Max. allowable input: 30Vrms or 60V DC (the  
1/80 of measurement range (using 12-bit A/D conversion; installed in the 8841/8842)  
1MS/s (simultaneous sampling in 2 channels)  
Measurement resolution  
Highest sampling rate  
Accuracy  
maximum voltage that can be applied across input pins without damage)  
DC amplitude: ±0.4% of full scale, zero position: ±0.1% of full scale (after zero adjustment)  
DC to 400kHz ±3dB, with AC coupling: 7Hz to 400kHz ±3dB  
DC, GND, AC  
Voltage: 0.5mV to 2V/DIV, 12 ranges  
Current: 5mA to 100A/DIV, 10 ranges, using current sensor (powered  
from the 8940, max. 4 sensors total)  
DC amplitude accuracy: ±0.4% of full scale, zero position ±0.15%  
of full scale (current measurement accuracy dependent on sensor accuracy/  
Frequency characteristics  
Input coupling  
Max. allowable input  
400V DC (the maximum voltage that can be applied across input pins without damage)  
Measurement  
range  
characteristics)  
Frequency characteristics: DC to 400kHz ±3dB  
Full scale: 20DIV, Max. allowable input: 30Vrms or 60V DC (the  
maximum voltage that can be applied across input pins without damage)  
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80 % rh after 30 minutes of warm-up  
FFT ANALOG UNIT 8938  
time and zero-adjust; accuracy guaranteed for 1 year)  
Number of channels: 2, for voltage measurement  
Measurement functions  
Anti-aliasing filter  
Other functions  
1/80 of measurement range (installed in the 8841/8842, excluding current range when using 9279)  
Measurement resolution  
Highest sampling rate  
Integrated filter for suppressing aliasing distortion caused by FFT  
processing (automatic cutoff frequency setting/OFF)  
1MS/s (simultaneous sampling in 2 channels), (frequency/duty ratio  
measurement: 1.125μs cycle)  
Other specifications same as the ANALOG UNIT 8936  
Voltage input pull-up: ON (10kΩ)/OFF, input coupling: DC, GND, AC  
(voltage/current), DC (others), low-pass filter: 5Hz/500Hz/5kHz/100kHz  
Other functions