3
Temperature/Analog, all inputs isolated  
- Measurement Functions -  
Universal isolated temperature, voltage and pulse inputs  
Universal isolated temperature, voltage and pulse inputs,  
Measurement parameters  
Voltage  
Ranges  
100mV f.s.  
Range of Measurements Finest Resolution  
Universal measurement inputs, voltage, temperature  
(thermocouple and Pt inputs*1) and humidity*1,2 can be  
selected for each channel. In addition, four input channels  
are provided for measuring pulse inputs (totalization/  
rotation count) simultaneously with voltage, temperature  
and humidity. In addition to channel-to-channel input  
isolation, the PC connection interface is completely  
isolated from the measurement terminals. Shock hazard  
is minimized even when thermocouples and voltage  
inputs are measured at the same time.  
-100mV to +100mV  
-1V to +1V  
5 V  
μ
1V f.s.  
50 V  
μ
10V f.s.  
-10V to +10V  
-60V to +60V  
1V to 5V  
500 V  
μ
100V f.s.  
5mV  
1 − 5V f.s.  
100°C f.s.  
500°C f.s.  
2000°C f.s.  
50,000c f.s.  
500,000c f.s.  
5Mc f.s.  
500 V  
μ
-100°C to 100°C  
-200°C to 500°C  
-200°C to 2000°C  
0.01°C  
0.1°C  
0.5°C  
Temperature  
Thermocouples: K, E, J, T,  
N, W (WRe5-26), R, S, B  
0
to 50,000 counts  
to 500,000 counts  
to 5M counts  
1 count  
0
10 counts  
0
100 counts  
2,000 counts  
Pulse Accumulation  
(Maximum rated voltage above ground is 60 V DC.)  
100Mc f.s.  
2,500Mc f.s.  
5,000/n (r/s) f.s.  
100%rh f.s.  
0
to 100M counts  
1
*
Pt and humidity measurement inputs are supported only by the 8420-51 and  
8421-51.  
Requires the HUMIDITY SENSOR 9653 (both optional).  
0
to 2,500M counts  
50,000 counts  
1
1
1
2
*
*
*
0 to 5,000/n (r/s)  
1/n (r/s)  
*
Rotation  
Humidity  
5.0 to 95.0%rh  
0.1%rh  
*1 n = pulses per rotation (1 to 1,000)  
Sample multiple channels at high speeds  
All input channels are quickly scanned, measured and stored within 100 ms (200 ms or more with channels 17 to 32 in Model  
, and within  
8422-51  
about 5 seconds for mixed humidity measurements). As stand-alone instruments, Model 8420-51 provides 4 pulse input channels, plus 8 temperature/  
voltage channels, Model 8421-51 provides 4 pulse input channels, plus 16 temperature/voltage channels, and Model 8422-51 provides 4 pulse input  
channels, plus 32 temperature/voltage channels. The 32MB of internal memory records about 16.77 million data points.  
Real-Time Save to High-Capacity Memory Card  
Measurement data can be automatically saved to a PC  
Card. Binary (real-time) and text (post-measurement)  
formats can be selected. High-capacity Flash ATA  
cards up to 1GB can be used for continuous long-term  
recording. Choose binary in normal use.This recording  
method is linked to writing measurements in real time.  
The supplied Wv Wave Viewer software can convert  
the data into text format on a PC.  
Recording  
intervals  
64MB  
64MB  
64MB  
64MB  
64MB  
(using 1 channel)  
(using 4 channels)  
(using 8 channels)  
(using 16 channels) (using 32 channels)  
100ms  
200ms  
500ms  
1s  
36 days  
72 days  
9 days 4 days, 12 hours  
18 days days  
45 days 22 days, 12 hours  
2 days, 6 hours  
4 days, 12 hours  
11 days, 6 hours  
- NA -  
9
2 days, 6 hours  
5 days, 15 hours  
11 days, 6 hours  
180 days  
360 days  
90 days  
180 days  
45 days 22 days, 12 hours  
90 days  
225 days 112 days, 12 hours 56 days, 6 hours  
2s  
1 year, 355 days  
4 years, 340 days  
45 days 22 days, 12 hours  
5s  
1 year, 85 days  
10s  
9 years, 315 days 2 years, 170 days  
1 year, 85 days  
225 days  
1 year, 85 days  
1 year, 310 days  
112 days, 12 hours  
225 days  
20s  
- omitted -  
- omitted -  
- omitted -  
- omitted -  
- omitted -  
4 years, 340 days 2 years, 170 days  
7 years, 145 days 3 years, 255 days  
30s  
1 year, 85 days  
1 year, 310 days  
1min  
2min  
5min to 1hour  
- omitted -  
- omitted -  
- omitted -  
7 years, 145 days 3 years, 255 days  
- omitted -  
- omitted -  
7 years, 145 days 3 years, 255 days  
- omitted - - omitted -  
Real-Time Storage Recording Times with 64 MB Card (approximate times)  
Note: Recording times are calculated values, and cannot be guaranteed. For calculations, one year = 365 days.  
Calculated values resulting in extremely long periods are omitted.  
What happens if a power failure occurs while measuring ?  
We recommend using the real-time saving function of the MEMORY HiLOGGERs 8420-51, 8421-51 and  
8422-51 with a PC Card. This exclusive technology has been developed to preserve data as reliably as possible  
even in the event of a power failure by incorporating PC card technology with the know-how built into the  
MEMORY HiCORDER series recording instruments. When recording only to internal memory without using a  
card, stored data is retainedforabout tenminutes inthe event ofa powerfailure.  
Color LCD displayswaveformsandnumericalvaluessimultaneously, andallowsviewingearlierdatawhilemeasuring  
Scroll through the displayed graph while saving  
measurements in real time to PC Card to verify earlier  
measurements. You can also read the values at the  
movable cursors.  
Display of Earlier Waveforms  
Display of Current Measurements  
(The green bar at the bottom indicates (The green bar at the bottom indicates  
the relative location of the current the relative location of the current  
display in internal memory.)  
display in internal memory.)  
Standard Measurement Screen  
(Measured values appear numerically at the  
left, and plotted as a graph at the right together  
with the measured values at the cursors.)  