5
Flexible Measurement Signals  
Widen Scope forApplication  
Applications  
Evaluation of signal-dependent components  
Comparator setting screen with additional Δ% display  
Since any test signal can be selected, it is possible to  
measure the inductance of winding, floating capacitance,  
characteristics at operating frequency, and low  
frequency resistance components. The 3522-50 further  
allows inductance (L) and DC resistance (DCR) to be  
measured by the same unit.  
Measurement value  
display  
Deviation display  
Example of measuring signal dependence of coils  
For chokes, transformers, and other components  
with an inductive core, the values depend on the  
measurement signal. By varying the measurement current,  
measurements showing the signal dependence of the coil  
can be shown as a graph.  
Judgment standard value and upper and lower limit widths  
The screen at left shows an example of the Δ% setting;  
The screen at right shows an example of the % setting  
from current models. In either, the judgement range is a  
percentage of the reference values.  
The Δ% display is easy to interpret because the  
measurement value is displayed as a deviation.  
V mode  
V0  
The 3522-50 and 3532-50 provide three modes for  
selecting the measurement signal according to the  
component characteristics: open-circuit voltage (V),  
constant voltage (CV), or constant current (CC).  
Im =  
| R0 + Z|  
 
| Z |  
 
+ V0  
Vm =  
 
| R0 + Z |  
V mode : set V0  
CV mode : set V0 so that the voltage across the component is  
the CV value (VCV)  
CC mode : set V0 so that the current through the component is  
the CC value (ICC)  
CV mode  
CC mode  
VCV  
Im = ICC  
Im =  
 
 
| Z|  
ꢄꢇꢇ  
ꢈꢈꢈꢈꢈ  
Vm  
Im  
: voltage monitor value  
: current monitor value  
ꢆꢇꢉ  
Vm = VCV  
Vm = ICC × | Z|  
R0  
: output impedance (50 Ω constant)  
Evaluating battery characteristics by measuring  
the internal resistance  
By measuring the internal resistance of lead-acid or compact  
storage batteries, the state of deterioration of the battery, and its  
lifetime and characteristics can be determined.  
In particular, the 3522-50 provides low-  
frequency measurement from 1 mHz,  
allowing low frequency electrochemical  
impedance measurement, and other  
applications in basic chemical research.  
Series equivalent resistance Rs (Ω)  
Measurement values:  
Rs (DCR), Rs, |Z|, θ, etc.  
Measurement frequency:  
Frequency characteristics of a manganese battery  
(1 mHz to 100 kHz) [ cole-cole plot ]  
DC, 1 kHz fixed, and variable frequency  
Measurement signal:  
constant current (CC) mode  