6
Basic specifications
Measurement line
: Single phase 2 wire, single phase 3 wire, 3 phase 3 wire
(3V3A is possible), 3 phase 4 wire
Measurement item
: When using 9600, 9601, 9602 (optional)
Voltage, current, voltage/current peak, effective/reactive/apparent power,
power factor, phase, frequency, current/power integration, load rate, efficiency.
[Motor output (Pm) measured]
Measurement method
9603 (optional) external input unit required
Effective when 9603 ch A is torque (any of N/ m, mN/m, kN/m, kgf/m,
kgf/cm), and ch B is r/m
Display indication range : 0.1% to 130% of 9603 voltage range
(polarity not indicated)
: Digital calculation from measured voltage or pulse signal
When using the 9603 (optional)
Voltage, torque, r/min, frequency, motor output.
Calculation algorithm : Set to required format
When using the 9605 (optional)
Display indication range
: At the lowest range in the DC mode of Models 9600 & 9602: 0.2% to 130%
At the lowest range in the AC+DC mode of Models 9600 & 9602:
0.5% to 130%
At the 200mA range of Model 9601:
0.5% to 130%
At 0.1% to 130% of all other ranges.
All range is zero suppressed at less than lower % value.
Valid input range for voltage, current, and power is 0.5% to 110%
: 6.4 inch TFT color LCD (640
×
480 dot)
Display
Display resolution
: 99999 count
(except with integration)
, 9999999 count
(with integration)
Rectification method
: Switchable between RMS
(true root mean square value)
and MEAN
(average rectified RMS indication)
. When combined mode DC is
selected, its not possible to switch between them.
: 8 times/sec.
Display update rate
Combined mode
: DC, AC + DC, AC
(AC only when used in combination
Analog response time
: FAST (0.1 sec.), MID (0.8 sec.), SLOW (5.0 sec.)
Low pass filter
: OFF / 500Hz / 5kHz / 300kHz (-3dB) For 9601, 5k
/ 300kHz not available.
Polarity detection regulation filter
: OFF / 200Hz (-3dB)
Analog output
: Voltage / current / active power
DC ±5V f.s (1000V range is DC ±3.333 V f.s.)
Monitor output
: Voltage / current: 1Vrms f.s.
(1000V range is 0.6667 Vrms f.s.)
(Time required for stabilization to within ±1% when input is changed
from 0% to 90%, or 100% to 10% of range.)
with 9601 or 9602 + AC clamp-on sensor)
Harmonic, waveform, voltage fluctuation / flicker measurement function.
[Efficiency measurement]
Calculable factors : Maximum of 3 formats
Calculated items : P for each input unit, Pm when combined with 9603
(Items measured with 9605 not allowed)
[D/A output]
Number of channels
Output impedance
Output items
Output voltage
Output update rate
:
:
:
:
:
8ch (12 bit D/A converter polarity + 11bit)
100Ω±5%
Outputs 8 arbitrarily selected items
DC±5V/f.s.
16 times/sec
[FDD]
Compatible media
: 3.5-inch, 2HD (1.2MB/1.44MB)
: MS-DOS
Format
: Function for saving and loading settings, and saving
Functions
measured values. Function for outputting measured values.
Floppy disk formatting function, file renaming and deleting
[Interface]
GP-IB
RS-232C
Functions
Control signal level
Scaling
Averaging
: Conforms to IEEE-488.1 1987, with reference to IEEE-488.2 1987
: Start-stop synchronous, with baud rate of 2400 or 9600 bits/sec
: Integration start / stop control, Integration data reset, External
A/D
(For display update when power meter display is in hold mode)
,
Manual print control, FD save control
: From 0 / 5V logic signal or short//open contact signal
[External Control]
[Voltage/ Current/ Active power measurement]
Measurement range
: See Page 5 specifications for individual input units
[Other functions]
: PT/CT ratio Set range 0.0001 to 10000
: Time average (set interval time, timer time, and average of
realtime control time)
Moving average
(number of samples: 8/16/32/64)
Exponential average
(attenuation factor: 8/16/32/64)
Multilingual display : Japanese/English screen display switching
Set time (all types) : Interval control time
(10 sec to 100 hours)
in 10 sec increments
Timer control time
(1 min to 10000 hours)
in 1 minute increments
Realtime control time, 1 minute increments
(When used in combination with FDD or printer, auto select increments)
[Integration measurement]
Number of measurements
: 64 times/sec
Measurement range
: 0 to±9999999 TAh / TWh
(integration time up to 10,000 hours)
[Power factor/ Phase angle measurement]
Measurement range
[Frequency measurement]
: -1.0000 (lead) to 0.0000 to 1.0000 (lag)
-180˚(lead) to 0.00˚to 180.00˚(lag)
Number of channels : Max. 3ch
(selection of voltage or current for arbitrary channel)
Effective input range
: 0.5 Hz to 2 MHz
Measurement range : Auto / 50Hz/ 500Hz / 50 kHz / 2 MHz
[Harmonic / Flicker measurement]
9605 (optional) required
[Wave peak measurement]
Measurement items : Select either voltage or current for each unit
(Shows absolute value of max.)
Effective Input Range : Effective value of sine wave is within effective input permi-
ssible in the range
Per accuracy table on page 5
±1 dgt. with respect to calculation from measured value (U, I, P)
sum value is max. ±3dgt.
±1dgt
with respect to values calculated from measurements ( I, P)
Max.
±
3dgt
with respect to values calculated from measurements (U, I, P)
Max.
±
3dgt
with respect to values calculated from measurements (U, I, P)
±0.1% rdg. ±1 dgt.
(0˚C to 40˚C(32˚F to 104˚F), for sine wave
input between 10% to 130% of U/I range)
(at 10kHz to 100kHz)
Measurement item
: See Page 4 specifications for 9605
Measurement accuracy
(23˚C±5˚C
(73˚F ±9˚F)
, Less than 80% rh, warm up time greater than 1 hour, sine wave input, power factor = 1, in-phase voltage = 0)
:
V, A, W
Apparent / reactive
:
power
:
Integration
:
Power factor
:
Phase angle
:
Frequency
Wave peak
Motor output
Efficiency
ments of items substituted into algorithm
: ±1dgt for calculations of each measured value
: Max. ±7 dgt
with respect to values calculated from measure-
: ±1%
(at 0.5Hz to 1kHz)
, ±2%
(at 1kHz to 10kHz)
, ±10%
Location for use
: Indoors, altitude to 2000 m
Ambient use humidity
: Power meter, 0˚Cto 40˚C
(32˚F
to
104˚F)
, rh below 80%
(no condensation)
When using FDD / printer, 5˚C to 40˚C
(41˚F
to
104˚F)
, rh below 80%
Ambient storage humidity
: -10˚C to 50˚C
(14˚F
to
122˚F)
, rh below 80% and no condensation
Insulation resistance
: More than 100MΩ from DC500V
General Specifications
Thermal coefficient
: Within ±0.03% f.s/ ˚C
Effect of in-phase
: Within ±0.05% f.s
(1000 Vrms, 50/60Hz, between shorted voltage input terminals and case)
voltage
: ±0.15% f.s (power factor = 0)
Effect of power factor
: ±25ppm ±1dgt. (0 to 40˚C
(73˚F to 41˚F)
)
Actual time
: Display accuracy - ±0.2% f.s
D/A output
Analog output
: Display accuracy - ±0.2% f.s
Monitor output
: Display accuracy - ±0.2% f.s
(less than 100kHz)
Display accuracy - ±3dB
(100k to 1MHz)
: Safety
EN61010-1: 2001
: EMC
EN61326: 1997+A1: 1998+A2: 2001 class A
EN61000-3-2: 2000
EN61000-3-3: 1995+A1: 2001
: AC100V/120V/200V/230V
(switched automatically)
, 50/60Hz
Power supply
Maximum rated power
: 150VA max.
Dimensions, mass
: Approx 430 W
✕150
H
370 D mm, Approx13 kg
Certifications
(Approx 16.93ʺ(W ) 5.91ʺ (H)14.57ʺ (D), Approx 458.6 oz.)
(in configuration including 9600 5 6ch, 9603, 9604)
(Not including projections such as terminals, feet, and handles)
Greater than 50M( at DC500V between U/I input terminal and
unit case, between U/I input terminal and power supply plug
(for the 9600/9601), between U input terminal and clamp input
terminal, between U input terminal and unit case, and between
U input terminal and power supply plug (for 9602)
Withstand voltage
: AC5.55kV
between U/I terminal and unit case, between U/I
terminal and power supply plug (for 9600 and 9601), between U
(50/60 Hz, 1 minute)
terminal and clamp input terminal, between U terminal and unit
case, between U input terminal and power supply plug (for 9602)
Accessories
: Power cord 1, ground adapter (3P to 2P) 1,
connector 1