Non-operating Conditions
Temperature
–20 ꢀC to ꢁ60 ꢀC
15% to 95% RH
Humidity (at wet bulb temperature
≤45 ꢀC, without condensation)
Altitude
0 m to 4,572 m
Others
EMC
EN 55011(1991)/CISPR 11(1990) Group 1, Class A
EN 50082-1(1992)/IEC 61000-4-2(1995) 4 kV CD, 8 kV AD
EN 50082-1(1992)/IEC 61000-4-3(1995) 3 V/m, 27 MHz to 1 GHz
EN 50082-1(1992)/IEC 61000-4-4(1995) 0,5 kV Signal Line, 1 kV
Power Line
EN 61000-3-2(1995)/IEC 61000-3-2(1995)
EN 61000-3-3(1995)/IEC 61000-3-3(1994)z
Safety
EN 61010-1(1993) ꢁAmd2(1995)/IEC61010-1(1990)
ꢁAm1(1992) ꢁAm2(1995)
CSA-C22.2 N0.1010.1-92
Power requirement
90 V to 132 V, or 198 V to 264 V (automatically switched),
47 Hz to 63 Hz, 300 VA max.
Weight
25 kg (SPC)
Dimensions
See Figures 10 through 12.
28