Measurement Accuracy
Conditions of Accuracy Specifications
Temperature
Four-terminal pair port of the
4294A’s front panel
23 ±5 °C
Beyond 23 ±5 °C, the measurement accuracy is twice as bad
as described.
Within ±5 °C from the adapter setup temperature.
Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
Within ±5 °C from the adapter setup temperature.
Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
Within ±5 °C from the adapter setup temperature.
Measurement accuracy applies when the adapter setup is per-
formed at 23 ±5 °C. When the adapter setup is performed beyond
23 ±5 °C, the measurement accuracy is twice as bad as described.
5
7-mm port of the 42942A
Terminal Adapter
3.5-mm port of the 42941A
Impedance Probe
Measurement terminal of the
16048G or 16048H
Measurement bandwidth
Measurement Accuracy
|Z|, |Y| accuracy
±E [%] (see Equation 1 on page 10, Equation 2 on page 12,
Equation 3 on page 14)
±E /100 [rad]
θ
accuracy
L, C, X, B accuracy
at
D
x
≤0.1
at
D
x
>0.1
R accuracy
±E [%]
±E X
1+
`b
D
x
2
[%]
8