Trigger Function
Trigger type
Continuous, Single, Number of groups
Trigger source
Internal (Free run), External (BNC connector input), GPIB or LAN,
Manual (Front key)
Trigger event type
Point trigger, Sweep trigger
Measurement Bandwidth/Averaging
Measurement bandwidth
Range
Averaging
Type
1 (Fast) to 5 (Precise), 5 steps
Sweep-to-sweep averaging, Point averaging
1 to 256 (integer)
Averaging factor
Adapter Setup
Adapter Selection
NONE
No adapter (the 16047E, etc. direct connection type test fixture is
connected)
4TP 1M
Four-terminal pair 1 m (16048G)
Four-terminal pair 2 m (16048H)
Terminal Adapter (42942A)
Impedance Probe (42941A)
4TP 2M
7-mm 42942A
PROBE 42941A
Calibration
Calibration
User calibration
Calibration performed with user-defined calibration kit (OPEN,
SHORT, LOAD)
Port extension
Compensation performed when the measurement terminal is
expanded from the 7-mm connector of the 42942A Terminal
Adapter or the 3.5-mm connector of the 42941A Impedance
Probe. Enter electrical length or delay time for the extension.
Fixture compensation
Calibration points
Compensation performed at the device contacts of the test fix-
ture using OPEN, SHORT, LOAD.
Fixed points, or User points determined by sweep setups
7