Measurement Accuracy  
Conditions of Accuracy Specifications  
Temperature  
Four-terminal pair port of the  
4294A’s front panel  
23 ± 5 ꢀC  
Beyond 23 ± 5 ꢀC, the measurement accuracy is twice as bad  
as described.  
7-mm port of the 42942A  
Terminal Adapter  
Within ± 5 ꢀC from the adapter setup temperature.  
Measurement accuracy applies when the adapter setup is per-  
formed at 23 ± 5 ꢀC. When the adapter setup is performed beyond  
23 ± 5 ꢀC, the measurement accuracy is twice as bad as described.  
3.5-mm port of the 42941A  
Impedance Probe  
Within ± 5 ꢀC from the adapter setup temperature.  
Measurement accuracy applies when the adapter setup is per-  
formed at 23 ± 5 ꢀC. When the adapter setup is performed beyond  
23 ± 5 ꢀC, the measurement accuracy is twice as bad as described.  
Measurement terminal of the  
16048G or 16048H  
Within ± 5 ꢀC from the adapter setup temperature.  
Measurement accuracy applies when the adapter setup is per-  
formed at 23 ± 5 ꢀC. When the adapter setup is performed beyond  
23 ± 5 ꢀC, the measurement accuracy is twice as bad as described.  
Measurement bandwidth  
5
Measurement Accuracy  
|Z|, |Y| accuracy  
± E ±%] (see Equation 1 on page 10, Equation 2 on page 12,  
Equation 3 on page 14)  
θ accuracy  
± E /100 ±rad]  
L, C, X, B accuracy  
at Dx 0.1  
± E ±%]  
at Dx >0.1  
± E X 1 ꢁ Dx2 ±%]  
`b  
R accuracy  
8