Agilent 35670A Specifications  
Instrument specifications apply after 15 minutes warm-up  
and within 2 hours of the last self-calibration. When the  
internal cooling fan has been turned OFF, specifications  
apply within 5 minutes of the last self-calibration. All speci-  
fications are with 400 line frequency resolution and with  
anti-alias filters enabled unless stated otherwise.  
Accuracy  
Frequency  
±30 ppm (.003ꢀ)  
Maximum range**  
Single channel ampltude  
1 channel mode  
102.4 kHz,  
Absolute amplitude accuracy (FFT)  
(A combination of full scale accuracy, full scale flatness,  
and amplitude linearity.)  
51.2 kHz (opt AY6*)  
51.2 kHz  
2 channel mode  
4 channel mode (Option AY6 only) 25.6 kHz  
Spans  
1 channel mode  
2 channel mode  
4 channel mode (Option AY6 only) 97.7 mHz to 25.6 kHz  
Minimimum resolution  
1 channel mode  
±2.92ꢀ (0.25 dB) of reading  
±0.025ꢀ of full scale  
195.3 mHz to 102.4 kHz  
97.7 mHz to 51.2 kHz  
FFT full scale accuracy at 1 kHz (0 dBfs)  
±0.15 dB (1.74ꢀ)  
FFT full scale flatness (0 dBfs) relative to 1 kHz  
±0.2 dB (2.33ꢀ)  
122 µHz (1600 line  
display)  
61 µHz (1600 line  
display)  
FFT amplitude linearity at 1 kHz measured on +27 dBVrms  
range with time avg, 0 to -80 dBfs  
±0.58ꢀ (0.05 dB) of reading  
2 channel mode  
±0.025ꢀ of full scale  
Amplitude resolution  
4 channel mode (Option AY6 only) 122 µHz (800 line  
display)  
Maximum real-time bandwidth  
FFT span for continuous data acquistion)  
(Preset, fast averaging)  
(16 bits less 2 dB over-range) with averaging 0.0019ꢀ of  
full scale (typical)  
Residual DC response (FFT mode)  
Frequency display (excludes A-weight filter)  
<-30 dBfs or 0.5 mVdc  
1 channel mode  
2 channel mode  
25.6 kHz  
12.8 kHz  
FFT dynamic range  
4 channel mode (Option AY6 only) 6.4 kHz  
Spurious free dynamic range  
(Includes spurs, harmonic distortion, intermodulation  
distortion, alias products). Excludes alias responses at  
extremes of span.  
Measurement rate  
(Typical) (Preset, fast averaging)  
1 channel mode  
2 channel mode  
≥ 70 averages/sec  
≥ 33 averages/sec  
Source impedence = 50 Ω.  
4 channel mode (Option AY6 only) ≥ 15 averages/sec  
Display update rate  
800 line display.  
90 dB typical (<-80 dBfs)  
Typical (Preset, fast average off) ≥ 5 updates/Sec  
*
Option AY6 single channel maximum range extends to 102.4 kHz  
without anti-alias filter protection.  
Maximum  
≥ 9 updates/Sec  
(Preset, fast average off, single channel, single display,  
undisplayed trace displays set to data registers)  
** Show all lines mode allows display of up to 131.1, 65.5 and 32.7  
kHz respectively. Amplitudes accuracy is unspecified and not alias  
protected.  