Maximum input voltage
30 VRMS, 50 Vdc
Level Ranges
0 - 100 mV to 0 - 100 V RMS in a 1, 3, 10 sequence
Digital readout also in mW (user selectable)
Resolution
1 mV or 1% of reading
Indication
3 digits and bar-chart
Accuracy AC
±3% ±3 mV ± resolution up to 30 V RMS
Accuracy DC
±1% +50 mV up to 40 V
Resolution
0.1% distortion
Indication
3 digits and bar-charts
Accuracy
±1 dB of reading ± 0.5% distortion
Sensitivity
50 mV (100 mV for 1% distortion) reading suppressed if audio voltage
is less than 5 mV
AUDIO S/N METER
Range
0 to 30 dB and 0 to 100 dB
Resolution
0.1 dB
Indication
3 digits and bar-chart
Accuracy
±1 dB
Sensitivity
50 mV (100 mV for 40 dB S/N) reading suppressed if audio voltage is
less than 5 mV
AUDIO FREQUENCY METER
Frequency Range
20 Hz to 50 kHz
Resolution
0.1 Hz, less than 10 kHz
1 Hz, at 10 kHz and above
Indication
5 digits
Accuracy
As frequency standard ± 1 digit ± resolution
Sensitivity
50 mV
AUDIO OSCILLOSCOPE
Operating Modes
Single or repetitive sweep
Frequency Range
DC to 50 kHz, 3 Hz to 50 kHz AC coupled
Voltage Range
10 mV to 20 V per division in a 1, 2, 5 sequence
Voltage Accuracy
±5% of full scale
FM Ranges
±75, 30, 15, 6, 3 and 1.5 kHz deviation full scale, ±10% accuracy
AM Ranges
20, 10 and 5% per division, ±10% accuracy
Timebase
50
µ
s/div to 5 s/div in a 1, 2, 5 sequence
Graticule
10 Horizontal by 6 Vertical divisions
Special features
Built in antialiasing circuitry and variable decode trigger level
AUDIO SINAD METER
Frequency
1 kHz
Range
0 to 18 dB and 0 to 50 dB
Resolution
0.1 dB
Indication
3 digits and bar-charts
Accuracy
± 1 dB
Sensitivity
50 mV (100 mV for 40 dB SINAD) reading suppressed if audio voltage
is less than 5 mV
AUDIO DISTORTION METER
Frequency
1 kHz
Range
0 to 10% and 0 to 30%
AUDIO BAR CHARTS
Bar-chart Displays
AF Voltage, SINAD, Distortion, S/N
For the very latest specifications visit
www.aeroflex.com